Severe dialyzer dysfunction undetectable by standard reprocessing validation tests
Authors:J A Delmez  C A Weerts  P D Hasamear  D W Windus
Affiliation:Department of Medicine, Washington University School of Medicine, St. Louis, Missouri.
Abstract:It is generally accepted that careful monitoring of total cell volume and ultrafiltration rates will ensure adequate function of reprocessed dialyzers. During routine urea kinetic measurements we noted that the percent of patients with clearances less than 200 ml/min increased from 5% to 48% despite adherence to these validation tests. As these patients did not have evidence of recirculation in the vascular access, possible causes of dialyzer dysfunction were investigated. Injection of methylene blue into the dialysate port revealed non-uniform flow of dialysate in dialyzers from patients with markedly reduced clearances. In vitro studies of dialyzers subjected to sequential daily reprocessing, without patient exposure, demonstrated that in vitro clearances declined in one lot but not another. The initial clearances of 218 +/- 4 ml/min fell progressively to 112 +/- 18 (P less than 0.001) after 15 reuses. No effects of reprocessing were found in a different lot (230 +/- 2 vs. 226 +/- 4 ml/min). Soaking the dialyzers from the affected lot in either the disinfectant or dialysate solution caused a decline in the clearances which was less than that of serial reuse. Although the magnitude of the problem of dialyzer malfunction with reuse is unknown, careful attention to dialyzer function is warranted in patients treated with reprocessed dialyzers.