首页 | 官方网站   微博 | 高级检索  
     


Thermal lens spectrometric determination of hexavalent chromium
Authors:Mateja ikovec  Mladen Franko  Federico G Cruz  Sidney A Katz
Affiliation:Mateja Image ikovec, Mladen Franko, Federico G. Cruz,Sidney A. Katz,
Abstract:The applicability of thermal lens spectrometry (TLS) for quantification and routine determination of hexavalent chromium was investigated by using a collinear dual beam thermal lens spectrometer. In aqueous solutions the LOD of 0.1 μg l?1 was obtained for Cr(VI) by using 160 mW laser power. The performance of the technique was verified by the determination of hexavalent chromium in standard reference water samples (NIST SRM 1643a and NIST SRM 1643c) and comparing the results for Cr(VI) in CCA (Cr, Cu, As)-treated timber extracts to concentrations obtained by atomic absorption spectrometry (AAS). Good agreement between the TLS results and reported values for Cr(VI) in SRMs as well as AAS results for Cr(VI) in CCA-treated timbers confirmed that TLS is a reliable and accurate analytical technique applicable for the determination of Cr(VI) in aqueous solutions at concentration levels 0.5–100 μg l?1.
Keywords:Chromium  Thermal lens spectrometry  Timber extracts  Environmental analysis  Atomic absorption spectrometry  Timber extracts
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23