首页 | 官方网站   微博 | 高级检索  
     


Measurement of thermal expansivity of low-expansion glasses by interferometric methods: Results of an interlaboratory comparison
Authors:H. Imai  M. Okaji  T. Kishii  H. Sagara  H. Aikawa  R. Kato
Affiliation:(1) National Research Laboratory of Metrology, 1-1-4 Umezono, Tsukuba, 305 Ibaraki, Japan;(2) Toshiba Glass Corporation, 3583-5 Kawajiri, Yoshida-machi, Kanbara-gun, 421-03 Sizuoka, Japan;(3) Hoya Corporation, 572 Miyazawa-cho, Akishima-shi, 196 Tokyo, Japan;(4) Shinku-Riko Inc., 300 Hakusan-cho, Midori-ku, 226 Yokohama, Japan
Abstract:An interlaboratory testing program on the measurement of the linear thermal expansion coefficient of low-expansion glasses has been carried out. Three different types of interferometric dilatometers, each located at three different organizations, and two kinds of low-expansion glass materials were selected for the experiments. As a result of the comparison, a reasonable agreement among the different measuring instruments was confirmed, and it was determined that the thermal expansion coefficient for low-expansion glasses can be measured with an accuracy of ±4×10–8 K–1 by using commercially available interferometric dilatometers.
Keywords:coefficient of linear thermal expansion  dilatometry  interferometric dilatometer  low-expansion glasses  reference method  thermal expansion  standardization
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23