Disposable microfluidic blood cuvette for measuring hemoglobin concentration
Authors:Hyung Jin Kim  Jin Kug Son  Young Ho Seo  Byeong Hee Kim  Won Hee Lee  Ki Tae Park  Kieth Bong Nahm  Eui Yul Choi
Affiliation:1. Department of Mechanical and Mechatronics Engineering, Kangwon National University, Chuncheon, South Korea
2. Boditech Med Inc., G-tech Village, Chuncheon, South Korea
Abstract:This paper presents a new air-bubble free microfluidic blood cuvette for the measurement of hemoglobin concentration. The microfluidic blood cuvette was filled with blood samples by capillary force, and hemoglobin levels in the blood were determined by measuring absorbance at the wavelength of 530 nm. Two different microfluidic blood cuvettes with dual and single sidewall microchannels were investigated. The microfluidic blood cuvette was fabricated using a polymethyl methacrylate substrate and a dry film photoresist. During the blood-filling process, air was trapped in the dual-sided wall-type cuvettes, while no air trapping occurred in the single sidewall-type cuvettes. The sensitivity of the hemoglobin measurements was more linear in a 105 μm deep microchannel than in a 35 μm deep microchannel.