首页 | 官方网站   微博 | 高级检索  
     


Multicenter evaluation of the performance and clinical utility in longitudinal monitoring of the Bayer Immuno 1 complexed PSA assay.
Authors:W J Allard  C D Cheli  D L Morris  J Goldblatt  Y Pierre  L Kish  Y Chen  J Dai  R L Vessella  D W Chan  M K Schwartz  Z Zhou  K K Yeung
Affiliation:Bayer Corporation, Tarrytown, New York, USA. jeffrey.allard.b@bayer.com
Abstract:We conducted a multicenter evaluation of the analytical and clinical performance of the automated Bayer Immuno 1 complexed PSA (cPSA) assay, and compared assay performance to the Bayer Immuno 1 PSA assay. We sought to determine whether measurements of cPSA could be of clinical utility in the management of patients with prostate cancer. Results of the 10-day imprecision across three evaluation sites produced total CV < 2.50% and an analytical sensitivity of 0.02 microgram/L. There was an increased trend in clinical sensitivity for prostate cancer with increasing stage of disease (71-86%). Clinical specificity for patients with benign urogenital disease was 74.8%, and for other nonprostate diseases ranged from 91.1-100%. Retrospective serial monitoring of 155 patients with prostate cancer demonstrated concordance of cPSA measurements to clinical status for 97% of the patients analyzed. Results from the clinical studies using the Bayer Immuno 1 cPSA assay were comparable to results obtained with the Bayer Immuno 1 PSA assay. The Bayer Immuno 1 cPSA assay demonstrates analytical performance and clinical effectiveness in the management of prostate cancer patients during the course of disease and therapy.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号