宽波段单色仪多级谱高精度波长定标
引用本文:张子辉,王淑荣,黄煜,李博,于向阳,林冠宇.宽波段单色仪多级谱高精度波长定标[J].光谱学与光谱分析,2012,32(10):2870-2874.
作者姓名:张子辉  王淑荣  黄煜  李博  于向阳  林冠宇
作者单位:1. 中国科学院长春光学精密机械与物理研究所应用光学国家重点实验室,吉林 长春 130033
2. 中国科学院研究生院,北京 100049
基金项目:国家自然科学基金项目,吉林省青年科研基金项目
摘    要:通常利用单色仪输出的单色光对空间遥感光谱仪进行波长定标。提出以空间遥感光谱仪的置信度为标准,来评价宽波段单色仪高精度波长定标精度的方法。通过对仪器精度的分析,分别求出单色仪的波长的重复性误差和偏差。应用高压汞灯的本征谱和光栅衍射多级谱作为定标谱线,避免更换灯源带来的误差。通过粗细定标相结合的方法,缩短扫描时间,并且运用高斯拟合对波峰进行精确定位,缩小误差。最后利用高次拟合得到的关系式,测出单色仪波长精度,计算出空间遥感光谱仪定标的置信度。以1.5 M单色仪为例,单色仪在200~840 nm波段内波长精度±0.016 nm,则空间遥感光谱仪的波长精度达到±0.050 nm的置信度为99.82%。

关 键 词:光谱学  波长定标  寻峰  多级谱  高斯拟合  
收稿时间:2012-04-26

High-Precision Wavelength Calibration of Wide-Band Monochromator
ZHANG Zi-hui , WANG Shu-rong , HUANG Yu , LI Bo , YU Xiang-yang , LIN Guan-yu.High-Precision Wavelength Calibration of Wide-Band Monochromator[J].Spectroscopy and Spectral Analysis,2012,32(10):2870-2874.
Authors:ZHANG Zi-hui  WANG Shu-rong  HUANG Yu  LI Bo  YU Xiang-yang  LIN Guan-yu
Affiliation:1. State Key Laboratory of Applied Optics, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China2. Graduate University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:Usually the monochromator is used to output monochromatic light to calibrate the space remote sensing spectrometer. In the present paper, the confidence of space remote sensing spectrometer is used as a standard to evaluate the precision of wide-band monochromator wavelength calibration. Through analysis of the accuracy of the instrument, the monochromator wavelength repeatability error and deviations was obtained respectively. And the intrinsic spectrum of the high pressure mercury lamp and the grating diffraction was used as calibration lines to avoid the error caused by replacing the light source. Through the special method of wavelength calibration to shorten the scan time, the Gaussian fitting was used to look for peaks of wavelength to reduce error. Finally, the relationship derived from polynomial fitting to measure the exact wavelengths’ accuracy of the monochromator and calculate the calibration confidence of the space remote sensing spectrometer. Using this method, we can make wavelength accuracy of the 1.5 M monochromator with wavelength band from 200 to 840 nm to reach to ±0.016 nm, then the confidence of the space remote sensing spectrometer can reach to 99.82%.
Keywords:Spectroscopy  Wavelength calibration  Peak search  Multi-level spectrum  Gaussian fit  