首页 | 官方网站   微博 | 高级检索  
     


Gated fiber-optic-coupled detector for in vivo real-time radiation dosimetry
Authors:Justus Brian L  Falkenstein Paul  Huston Alan L  Plazas Maria C  Ning Holly  Miller Robert W
Affiliation:Optical Sciences Division, Naval Research Laboratory, Code 5611, 4555 Overlook Avenue, SW, Washington, DC 20375, USA. justus@nrl.navy.mil
Abstract:Gated detection of the output of a fiber-optic-coupled radiation dosimeter effectively eliminated the direct contribution of Cerenkov radiation to the signal. The radiation source was an external beam radiotherapy machine that provided pulses of 6-MeV x rays. Gated detection was used to discriminate the signal collected during the radiation pulses, including Cerenkov interference, from the signal collected between the radiation pulses due only to phosphorescence from the Cu(1+)-doped glass detector. Gated detection of the long-lived phosphorescence of the Cu(1+)-doped glass provided real-time dose measurements that were linear with the absorbed dose and that were accurate for all field sizes studied.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23