首页 | 官方网站   微博 | 高级检索  
     

基于热电偶的精密环境温场测量的关键问题研究
引用本文:李洪波,赵维谦,高思田,崔建军,丁健.基于热电偶的精密环境温场测量的关键问题研究[J].哈尔滨理工大学学报,2010,15(4):1-5.
作者姓名:李洪波  赵维谦  高思田  崔建军  丁健
作者单位:1. 北京理工大学,光电学院,北京,100081;中国航空工业总公司,北京长城计量测试技术研究所,北京,100095
2. 北京理工大学,光电学院,北京,100081
3. 中国计量科学研究院,北京,100013
基金项目:财政部公益行业专项基金 
摘    要:分析了热电偶温度测量不确定度、温场稳定度测量不确定度和温场均匀度测量不确定度,为精密环境的温场参数测量和控制提供理论保障.分析表明,热电偶测温不确定度与热电偶热电动势、参考端温度传感器的测量不确定度密切相关;温场均匀度测量不确定度与热电偶热电动势的测量不确定度密切相关;温场稳定度测量不确定度与热电偶热电动势和温差拟合函数的关系密切,在一次线性拟合的条件下取决于拟合函数的斜率.针对测量过程中存在的脉冲噪声和热电偶非线性的干扰,结合精密环境温度信号变化缓慢的特点,提出综合运用均值滤波和中值滤波处理热电偶热电动势测量数据去除测量中脉动噪声和热电偶非线性对测量结果的影响.

关 键 词:热电偶  测量不确定度  温场均匀度  温场稳定度  非线性  滤波

Research on Some Key of Thermocouple Used to Detect the Temperature of Precision Measurement Environment
LI Hong-bo,ZHAO Wei-qian,GAO Si-tian,CUI Jian-jun,DING Jian.Research on Some Key of Thermocouple Used to Detect the Temperature of Precision Measurement Environment[J].Journal of Harbin University of Science and Technology,2010,15(4):1-5.
Authors:LI Hong-bo  ZHAO Wei-qian  GAO Si-tian  CUI Jian-jun  DING Jian
Affiliation:1.School of Optoelectronics,Beijing Institute of Technology,Beijing 100081,China;2.Changchen Iustitute of Metrology and Measurement,Beijing 100095,China;3.National Institute of Metrology,Beijing 100013,China)
Abstract:Uncertainty of temperature measurement by thermocouple is analyzed,at same time,the uncertainty of stability and uniformity of temperature measurement are analyzed as well.The analysis brings forward the theoretical support for the parameters measurement and control of temperature field under precise environment.It shows the uncertainty of temperature measurement is closely related to the measurement uncertainty of both thermocouple EMF and the sensor temperature at reference-side.The measurement uncertainty of the uniformity of temperature is closely related to the uncertainty of EMF of the thermocouple,while the uncertainty of the stability of temperature is closely related to both EMF and the fitting function of temperature difference,which depends on the slope when the fitting function is a linear fitting.Taking the character that temperature change slowly in precise environment into account,the average filter combined with the median filter are used to remove or weaken the pulse noise and nonliner interference of the thermocouple during the EMF measurement.
Keywords:thermocouple  uncertainty  uniformity of temperature  stability of temperature  nonlinear  filtering
本文献已被 CNKI 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号