首页 | 官方网站   微博 | 高级检索  
     

标准铂电阻温度计自热效应对测量结果的影响
引用本文:任建平,孙建平,李婷,何佳融,曾佳旭.标准铂电阻温度计自热效应对测量结果的影响[J].计量学报,2021,42(5):589-594.
作者姓名:任建平  孙建平  李婷  何佳融  曾佳旭
作者单位:1.山西省检验检测中心(山西省标准计量技术研究院),山西 太原 030032
2.中国计量科学研究院,北京 100029
3.吉林省计量科学研究院,吉林 长春 130022
4.中国计量大学,浙江 杭州 310018
基金项目:国家重点研发项目(2017YFF0205904)
摘    要:为了研究自热效应对标准铂电阻温度计测量结果的影响,分别从定点法和比较法两方面开展研究。针对定点法,统计中国计量科学研究院近3年检定的标准铂电阻温度计数据,计算不同温区的铂电阻温度计自热效应修正前后的测量结果并进行对比,结果表明,在溯源标准铂电阻温度计时自热效应修正与否对测量结果的影响达到了1.5mK以上,最大达到了6mK。针对比较法,设计了基于恒温槽的自热测量方案并进行实验,对修正自热前后温度测量结果进行对比,结果表明,自热修正与否带来的误差依然达到了1.5mK。因此,在标准铂电阻温度计量值溯源和精密温度测量中,对测量结果进行自热修正是必要的。

关 键 词:计量学  标准铂电阻温度计  90国际温标  自热效应  
收稿时间:2020-01-03

Influence of Self-heating Effect of Standard Platinum Resistance Thermometer on Measurement Results
REN Jian-ping,SUN Jian-ping,LI Ting,HE Jia-rong,ZENG Jia-xu.Influence of Self-heating Effect of Standard Platinum Resistance Thermometer on Measurement Results[J].Acta Metrologica Sinica,2021,42(5):589-594.
Authors:REN Jian-ping  SUN Jian-ping  LI Ting  HE Jia-rong  ZENG Jia-xu
Affiliation:1. Shanxi Inspection and Testing Center(Shanxi Institute of Standardization and Metrology Technology),
Taiyuan, Shanxi 030032, China
2. National Institute of Metrology, Beijing 100029, China
3. Jilin Institute of Metrology and Research, Changchun, Jilin 130022, China
4. China Jiliang University, Hangzhou, Zhejiang 310018, China
Abstract:In order to study the effect of self-heating on the measurement results of SPRTs, the research was carried out from the fixed-point method and the comparative method. For the fixed-point method, based on the statistics of SPRT verified by the National Institute of Metrology of China in the past three years, the measurement results of SPRT in different temperature regions before and after self-heating correction were compared. The results showed that the effect of self-heating correction on the value traceability of SPRT was more than 1.5mK, and the maximum reached 6mK. For comparative method, a self-heating measurement scheme based on a constant temperature bath was designed and experimented, and the temperature measurement results before and after the self-heating correction were also compared. The results showed that the deviation caused by self-heating correction still reached 1.5mK. Therefore, it is necessary to correct the effect of self-heating in the precision temperature measurement and value traceability of SPRT.
Keywords:metrology  standard platinum resistance thermometer  ITS-90  self-heating effect  
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《计量学报》浏览原始摘要信息
点击此处可从《计量学报》下载全文