1.5 T MR加速器临床剂量调试
引用本文:李明辉,田源,张可,牛传猛,王宏凯,门阔,戴建荣.1.5 T MR加速器临床剂量调试[J].中华放射肿瘤学杂志,2022,31(1):29-34.
作者姓名:李明辉  田源  张可  牛传猛  王宏凯  门阔  戴建荣
作者单位:国家癌症中心/国家肿瘤临床医学研究中心/中国医学科学院北京协和医学院肿瘤医院放疗科,北京 100021
摘    要:目的:介绍1.5 T MR加速器的临床剂量调试方法和结果。方法:中国医学科学院肿瘤医院于2019年5月安装1台医科达Unity型1.5 T MR加速器,使用磁场兼容测量设备对其进行剂量调试。调试项目包括绝对剂量校准、数据采集和计划系统模型验证。结果:磁场下绝对剂量校准需要使用磁场修正因子修正,参考条件下的绝对剂量为87...

关 键 词:磁共振加速器  剂量调试  模型
收稿时间:2021-09-06

Clinical dosimetry commissioning of 1.5 T MR-linac
Li Minghui,Tian Yuan,Zhang Ke,Niu Chuanmeng,Wang Hongkai,Men Kuo,Dai Jianrong.Clinical dosimetry commissioning of 1.5 T MR-linac[J].Chinese Journal of Radiation Oncology,2022,31(1):29-34.
Authors:Li Minghui  Tian Yuan  Zhang Ke  Niu Chuanmeng  Wang Hongkai  Men Kuo  Dai Jianrong
Affiliation:Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China
Abstract:Objective To introduce the clinical dosimetry commissioning methods and results of the 1.5 T MR-linac. Methods In May, 2019, an Elekta Unity 1.5 T MR-linac was installed in Cancer Hospital, Chinese Academy of Medical Sciences and dosimetry commissioning was performed with magnetic field compatible measuring instruments. Commissioning items include absolute dose calibration, data acquisition and planning system model verification. Results Absolute dose calibration in magnetic field should be corrected by magnetic field correction factor. The standard output dose of Unity was 87 cGy. Gamma analysis (3%/2mm) was performed on the beam collection data and the planning system calculation data. The average pass rate of dose verification of standard field test cases was 96.41%, and the TG119 test case was 98.24%. The IROC end to end test case was 97.5%(7%/4mm). Conclusions The planning system model and the beam collection data have good consistency. The dose verification results of the standard field and TG119 test cases meet the general tolerance limit requirements of the AAPM TG218 report, and the verification results of the IROC end-to-end test cases meet the IROC center standards.
Keywords:Magnetic resonance accelerator  Dosimetry commissioning  Model