首页 | 官方网站   微博 | 高级检索  
     


Flow amperometric detection of indigo for enzyme-linked immunosorbent assays with use of screen-printed electrodes
Authors:Setsuko YajimaYuri Sonoyama  Ken SuzukiKeiichi Kimura
Affiliation:a Faculty of Systems Engineering, Wakayama University, Sakae-dani, Wakayama, 640-8510 Wakayama, Japan
b Tsukuba Research Laboratory, NOF Corporation, Toukoudai, Tsukuba, 300-2635 Ibaraki, Japan
Abstract:A simple and automated methodology for a sensitive electrochemical detection of enzyme immunoassays that employ alkaline phosphatase (AP) as label has been developed. A flow injection system with programmable pump, valve and cell functions, amperometric detection of indigo and screen-printed electrodes (SPEs) are responsible for the advantages of this methodology. Amperometric detection at a low potential of indigo, the product of the enzymatic hydrolysis of the substrate 3-indoxyl phosphate (IP), is combined with a flow injection system. This incorporates in the flow cell a disposable screen-printed board provided with a graphite working electrode. No electrode pretreatment is necessary to obtain reproducible signals. The system was applied to the determination by an enzyme-linked immunosorbent assays (ELISA) of pneumolysin (PLY), a toxin related to respiratory infections. Linear calibration curves for low and high concentration ranges were obtained. These were also performed in a proteic matrix and linearity was also obtained.
Keywords:Flow electrochemical enzyme immunoassay   Screen-printed electrodes   ELISA   Pneumolysin   3-Indoxyl phosphate   Alkaline phosphatase
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏