Two-dimensional accelerated MP-RAGE imaging with flexible linear reordering |
| |
Authors: | Daniel Brenner Rüdiger Stirnberg Eberhard D. Pracht Tony Stöcker |
| |
Affiliation: | 1. German Center for Neurodegenerative Diseases (DZNE), Ernst-Robert-Curtius-Stra?e 12, 53117, Bonn, Germany
|
| |
Abstract: | Object Implementation of an accelerated Magnetization Prepared RApid Gradient Echo (MP-RAGE) sequence for T1 weighted neuroimaging; exploiting modern MRI technologies to minimize scan time while preserving the image quality. Materials and methods A custom MP-RAGE sequence was implemented on a state-of-the-art 3T MR scanner equipped with a 32-channel receiver array head coil. The sequence utilized a shifted CAIPIRINHA k y –k z under-sampling pattern combined with elliptical scanning and a two-dimensional view ordering scheme to achieve high parallel imaging acceleration factors at maintained image contrast. Results It could be shown that MP-RAGE accelerated in two k-space directions outperforms single direction acceleration, which is the common practice with standard view ordering. Applying the CAIPIRINHA technique in conjunction with elliptical scanning further increased this benefit. Conclusion By combining MP-RAGE with CAIPIRINHA sampling and elliptical scanning, the scan time can be reduced from 4–5 min to 2–3 min with insignificant reduction in image quality. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|