首页 | 官方网站   微博 | 高级检索  
     


A proportional counter for low intensity heavy ion flow measurements
Authors:Yu G Teterev and G A Kononenko
Abstract:A proportional counter designed for heavy ion flux measurements is described. The working gas of the counter is air. The counter is composed of three parallel glass-fiber frames with wire electrodes coiled on them. Based on the counter, there has been developed a diagnostic product measuring the density distribution of accelerated ion flux for radioelectronic device testing. To prevent the dependence of measurement results on weather conditions (atmospheric pressure, temperature, humidity) as well as on the type and energy of ions, there has been applied an automatic correction methodology. In the process of the set-up operation it has been established, that the counter is characterized by a high radiation resource, resistance to extreme downloads, adequate performance of up to 106 ions/s, it is not sensitive either to neutrons or to γ-quanta, and it is convenient to use.
Keywords:
本文献已被 SpringerLink 等数据库收录!