首页 | 官方网站   微博 | 高级检索  
     

浅谈电视信号发生器PM5415的校准
引用本文:胡晴宇.浅谈电视信号发生器PM5415的校准[J].电子质量,2012(12):61-63.
作者姓名:胡晴宇
作者单位:江苏省电子信息产晶质量监督检验研究院,江苏无锡,214073
摘    要:根据信息产业部电子601计量站和信息产业部电子203计缱站起草的JJF1235—2010《电视视频信号发生器校准规范》,制式为PAL、NTSC、SECAM的模拟电视信号发生器都能用此校准规范进行校准,该文根据此校准规范埔VM700视频分析仪校准FLIJKE公司生产的璎号为PM5415的电视信号发生器。

关 键 词:电视信号发生器  校准规范  亮度电平  色度电平  相位

The Calibration of the PM5415 TV Signal Generator
Hu Qing-yu.The Calibration of the PM5415 TV Signal Generator[J].Electronics Quality,2012(12):61-63.
Authors:Hu Qing-yu
Affiliation:Hu Qing-yu(Jiangsu Electronic Information Products Quality Supervision and Isnpection Institute,Jiangsu Wuxi 214073)
Abstract:In this paper introduced that author used VM700 video analyzer to calibrate PM5415 TV Signal generator which was made by FLUKE Company.According to JJF1235-2010 《TV Video Signal Generator Calibration Specification》,which was drafted by the No. 601 & No. 203 Metrological Verification Station, member of the Ministry of Industry and Information Technology of PRC,that PAL,NTSC,SECAM system analog TV signal generator can be calibrated with this specification.
Keywords:TV Video Single Generator  Calibration Specification  Intensity Level  Chroma level  Phase