首页 | 官方网站   微博 | 高级检索  
     


Automatic seizure detection: improvements and evaluation.
Authors:J Gotman
Affiliation:Department of Neurology and Neurosurgery, McGill University, Montreal, Que., Canada.
Abstract:Improvements to an existing automatic seizure detection program are described. They are aimed at taking into account a larger temporal context and thus improving the specificity of the detections. Results were evaluated on 293 recordings from 49 patients, totaling 5303 h of 16-channel recording. They showed that 24% of the 244 seizures recorded were missed by the automatic detection; in 41% of the seizures, the patient alarm was not pressed but the computer made detections. The false detection rate was of the order of 1 false detection per hour of recording. Conclusions are: (1) automatic seizure detection must be used in conjunction with a patient alarm button since some seizures, having poorly defined EEG activity, are not detected; (2) the automatic detection allowed capture of many seizures, clinical and subclinical, for which the alarm was not pressed; (3) the low false detection rate indicates that lower detection threshold could be used, yielding better seizure detection.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号