光子辐射场配置电磁场研究磁共振引导放疗(MRIgRT)参考剂量磁场因子的可行性
引用本文:齐雅平,王志鹏,姚金涛,刘福斌,金孙均,黄骥,杨小元,张健,王坤. 光子辐射场配置电磁场研究磁共振引导放疗(MRIgRT)参考剂量磁场因子的可行性[J]. 中华放射医学与防护杂志, 2023, 43(4): 291-296
作者姓名:齐雅平  王志鹏  姚金涛  刘福斌  金孙均  黄骥  杨小元  张健  王坤
作者单位:中国计量科学研究院电离辐射计量科学研究所, 北京 100029;安徽理工大学机械工程学院, 淮南 232001
基金项目:国家重点研发计划资助(2021YFF0603600);基本科研业务经费(AKYCX2207)
摘    要:目的针对磁共振引导放疗(MRIgRT)射束参考剂量准确测量的需求, 研究强磁场对辐射剂量测量的影响, 验证医用加速器配置可变磁场测量磁场因子的可行性。方法利用常规医用电子直线加速器配置一对磁场强度高达1.5 T电磁铁所产生标称高压6 MV光子辐射场和可变磁场, 使用PTW30013和PTW31010电离室, 分别测试电离室轴与磁场方向夹角为0°、180°、90°和270°时0~1.5 T磁场强度范围内的电离室响应, 对测量结果进行分析, 计算磁场因子kB, M。结果电离室响应读数随磁场强度增加而增加, 并在1 T时达到峰值, 当磁场强度继续增加, 电离室读数随之降低。0.35 T磁场时, PTW31010电离室轴向与磁场方向垂直90°和平行0°下的电离室响应磁场因子kB, M分别为0.988 2±0.000 3和0.997 4±0.000 4, 后者与文献报道结果间差异为0.05%±0.04%。1.5 T磁场时, PTW30013电离室轴向与磁场垂直90°下kB, M为0.958 9±0.000 5,与文献值相差0.60%±0.05%。结论在6 MV医用加速器配置电磁铁所测量的电离室响...

关 键 词:磁共振引导放疗  磁场  电离室  磁场因子
收稿时间:2022-12-06

Feasibility study of integrating medical Linac with electromagnet to measure magnetic factor of reference dosimetry for MRIgRT
Qi Yaping,Wang Zhipeng,Yao Jintao,Liu Fubin,Jin Sunjun,Huang Ji,Yang Xiaoyuan,Zhang Jian,Wang Kun. Feasibility study of integrating medical Linac with electromagnet to measure magnetic factor of reference dosimetry for MRIgRT[J]. Chinese Journal of Radiological Medicine and Protection, 2023, 43(4): 291-296
Authors:Qi Yaping  Wang Zhipeng  Yao Jintao  Liu Fubin  Jin Sunjun  Huang Ji  Yang Xiaoyuan  Zhang Jian  Wang Kun
Affiliation:Division of Ionizing Radiation Metrology, National Institute of Metrology, Beijing 100029, China;College of Mechanical Engineering, Anhui University of Science and Technology, Huainan 232001, China
Abstract:Objective To study the influence of intensive magnet fields on radiation dose measurement, and to demonstrate the feasibility of measuring magnet field correction factor by a combination of medical linac with variable magnet fields in view of needing for accurate measurement of the doses from reference beam arising in MR image-guided radiotherapy.Methods A photon radiation field and a variable field with 6 MV nominal high voltage were produced by using conventional medical electron linear accelerator equipped with a pair of electromagnets with magnetic field strength up to 1.5 T. Both PTW30013 and PTW31010 ionization chambers were used to test the responses of ionization chambers under different magnetic field strengths at four orientations in which the angles between ionization chamber axis and magnetic field direction were 0°, 180°, 90° and 270°, respectively. The magnetic factors, kB,M was calculated and compared with the reported values in literature.Results The response of ionization chamber was proportional to the magnetic field strength before it reached to a peak around 1 T, and then fell down as the magnetic field continued to rise. When the magnetic field was 0.35 T, the magnetic factors of PTW31010 were 0.988 2±0.000 3 and 0.997 4±0.000 4 corresponding to 90° and 0° directions, the discrepancy between 0° scenario and literature was 0.05% ±0.04%. When the magnetic field reached 1.5 T, the magnetic factor of PTW30013 was 0.958 9±0.000 5 at the situation of 90°, which was 0.60% ±0.05% different from the literature value.Conclusions Conventional 6 MV medical accelerator equipped with electromagnet can be used to measure the magnetic field factor of reference dosimetry for MRIgRT.
Keywords:MRIgRT  Magnetic field  Ionization chamber  Magnetic factor