新型空腔式黑体辐射源设计及不确定度评定
引用本文:刘培,杨雪,王景辉,胡蝶,张鑫,徐标. 新型空腔式黑体辐射源设计及不确定度评定[J]. 计量学报, 2022, 43(2): 264-267. DOI: 10.3969/j.issn.1000-1158.2022.02.20
作者姓名:刘培  杨雪  王景辉  胡蝶  张鑫  徐标
作者单位:1. 广东省计量科学研究院,广东 广州 510405
2. 中国计量科学研究院,北京 100029
基金项目:广东省市场监督管理局科技计划项目
摘    要:为促进恒温槽内空腔黑体与载热工质之间的热交换,提出了一种附加被动扰流装置的黑体空腔辐射源的设计方法.通过新设计的空腔外壁微扰动结构,对载热工质施加干扰,使流体热边界层破坏并再生,利用热边界层初段传热强度高的原理强化传热.此设计既可以提高换热系数,又可以增加换热面积,减少空腔与载热工质达到热平衡所需的时间.对新型的耳温计...

关 键 词:计量学  黑体空腔辐射源  扰流结构  稳定时间  耳温计  不确定度评定
收稿时间:2021-08-25

Design and Uncertainty Evaluation for New Blackbody Source
LIU Pei,YANG Xue,WANG Jing-hui,HU Die,ZHANG Xin,XU Biao. Design and Uncertainty Evaluation for New Blackbody Source[J]. Acta Metrologica Sinica, 2022, 43(2): 264-267. DOI: 10.3969/j.issn.1000-1158.2022.02.20
Authors:LIU Pei  YANG Xue  WANG Jing-hui  HU Die  ZHANG Xin  XU Biao
Affiliation:1. Guangdong Provincial Institute of Metrology, Guangzhou, Guangdong 510405, China
2. National Institute of Metrology, Beijing 100029, China
Abstract:In order to enhance the heat transfer between the blackbody cavity and the heat carrier, a new design of cavity blackbody source with passive spoiler is proposed. With the newly designed micro-disturbance structure on the outer wall of the cavity, the thermal boundary layer of the working fluid is disturbed to destroy and regenerate, so the heat transfer is enhanced by the principle of high heat transfer intensity in the initial section of the thermal boundary layer. The new design can not only improve the heat transfer coefficient, but also increase the heat transfer area, which can reduce the time required for the cavity and the heat carrier to achieve thermal balance. The experimental study and uncertainty evaluation of the new blackbody radiation source of the ear thermometer are carried out. The results show that the expanded uncertainty of the new design cavity blackbody is U=0.05℃(k=2), and the equilibration time can be reduced by more than 15%.
Keywords:metrology  blackbody cavity  spoiler  equilibration time  ear thermometer  uncertainty evaluation  