首页 | 官方网站   微博 | 高级检索  
     

巯基化合物的电位分析法研究
引用本文:曾百肇,任健敏.巯基化合物的电位分析法研究[J].分析化学,1994,22(2):146-149.
作者姓名:曾百肇  任健敏
作者单位:武汉大学化学系!武汉430072(曾百肇,周性尧),江西农业大学基础部(任健敏)
摘    要:本文对用直接电位法测试巯基化合物的可行性进行了研究。发现由硫化银,硫化汞等金属难溶盐制成的膜电极对6-巯基嘌呤,高胱氨酸,1,4-二巯基苏糖经合物可产生响应,其斜率为30-110mV/pC,电极响应时间,则为十几秒至十几分钏,测试结果具有较好的重现性。因此,该方法可望得到应用。文中还对有关响应机理进行了探讨。

关 键 词:离子选择电极  巯基化合物  电位分析

Potentiometric Determination of Some Mercapto-containing Compounds
Zeng Baizhao,Ren Jianmin. Zhou Xingyao.Potentiometric Determination of Some Mercapto-containing Compounds[J].Chinese Journal of Analytical Chemistry,1994,22(2):146-149.
Authors:Zeng Baizhao  Ren Jianmin Zhou Xingyao
Abstract:The feasibility of the determination of some mercapto-containing compounds by ion-selective electrodes was investigated. It was found that several kinds of electrode, such as Ag_2S. HgS electrode etc. ,can respond to 6-mercaptopurine, 1, 4-dithiothreitol, homocys-tine etc. The response slope was about 30 to 110 mV/pC, response time was about 10 s to 10 min. Meanwhile, the electrodes exhibited good reproducibility. The response mechanism was also discussed.
Keywords:Ion-selective electrode  Mercapto-containing compounds  Potentiometric determination  
本文献已被 CNKI 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23