Ultra fast electron beam X-ray computed tomography for two-phase flow measurement
Authors:F. Fischer  U. Hampel
Affiliation:Institute of Safety Research, Forschungszentrum Dresden-Rossendorf e.V., POB 51 01 19, 01314 Dresden, Germany
Abstract:Electron beam X-ray CT is a new technique for a fast measurement of multiphase flows with frame rates of 1000 images per second and more. It gives, in principle, quantitatively accurate images of the flow at high spatial resolution and it is non-intrusive since moderately radiation absorbing vessel walls can be penetrated by X-rays. However, on the road to a technical realisation of such a technique within a computed tomography system many problems have to be solved. As a first prototype for scientific flow measurement studies, we devised and built a fast scanned electron beam X-ray tomography scanner. The scanner consists of an electron beam unit that can be operated at up to 150 kV acceleration voltage and up to 65 mA electron beam current, with the required electron optics for beam adjustment, beam focusing and beam deflection unit and a fast circular CZT detector comprising 240 elements of 1.5 mm × 1.5 mm × 1.5 mm active pixel area. X-ray radiation is produced on a circular water cooled tungsten target. The CT system achieves up to 7000 frames per second with a spatial resolution of 1 mm. First two-phase flow experiments have been carried out on gas-water flows in bubble columns. Further, a vertical test section made of titanium alloy has been installed at the TOPFLOW facility and will be used in future to study the evolution of two-phase gas-water pipe flow at high pressures and temperatures.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号