首页 | 官方网站   微博 | 高级检索  
     

HPLC-ELSD法测定生物柴油中的游离甘油
引用本文:李蓉,王武生,马晓迅,陈斌.HPLC-ELSD法测定生物柴油中的游离甘油[J].分析测试学报,2011,30(10):1167-1170.
作者姓名:李蓉  王武生  马晓迅  陈斌
作者单位:西北大学,化工学院,陕北能源先进化工利用技术教育部工程研究中心/陕西省洁净煤转化工程技术研究中心,陕西,西安,710069
基金项目:陕西省科技攻关计划项目(2009K10-02);陕西省自然科学基金项目(2011JM2013);陕西省教育厅专项科研计划项目(09JK735);西北大学科研启动基金项目(PR10028)
摘    要:采用液液萃取及高效液相色谱-蒸发光散射检测技术,建立了生物柴油中游离甘油含量的测定方法.待测生物柴油经乙腈-水(75:25)抽提并离心分离后,取部分下层水相进行分析.优化的色谱条件为:色谱柱Inertsil NH2 (250 nun ×4.6 mm,5μm );流动相为乙腈-水(75:25);流速1.0 mL/min;...

关 键 词:高效液相色谱法  蒸发光散射检测器  生物柴油  甘油

Determination on Content of Free Glycerol in Biodiesel by HPLC-ELSD
LI Rong,WANG Wu-sheng,MA Xiao-xun,CHEN Bin.Determination on Content of Free Glycerol in Biodiesel by HPLC-ELSD[J].Journal of Instrumental Analysis,2011,30(10):1167-1170.
Authors:LI Rong  WANG Wu-sheng  MA Xiao-xun  CHEN Bin
Affiliation:LI Rong,WANG Wu-sheng,MA Xiao-xun,CHEN Bin (Chemical Engineering Research Center of the Ministry of Education for Advanced Use Technology of Shanbei Energy,Shaanxi Research Center of Engineering Technology for Clean Coal Conversion,School of Chemical Engineering,Northwest University,Xi'an 710069,China)
Abstract:An analytical method was established for the determination of free glycerol in biodiesel by using liquid-liquid extraction technique and high performance liquid chromatography(HPLC) with evaporative light-scattering detector(ELSD).The biodiesel sample was extracted with acetonitrile-water(75∶25),followed by centrifugation.The obtained aqueous layer was analyzed by HPLC-ELSD.The testing conditions of HPLC-ELSD were optimized and the optimal conditions were as follows:separation column:Inertsil NH2 column(250 mm×4.6 mm,5 μm),mobile phase:mixture of acetonitrile and water(75∶25),flow rate:1.0 mL/min,injection volume:30 μL,drift tube temperature:28 ℃,gas pressure of nebulizer:360 kPa.The result indicated that,under the optimal conditions,the calibration curve was linear over the content of free glycerol in biodiesel in the range of 3.73-302.0 mg/kg with a correlation coefficient of 0.999 8,and the limits of detection(S/N≥3)were 2.1 mg/kg.The spiked recoveries were in the range of 98%-101% with relative standard deviations(RSDs) of 0.59%-1.8%.The method was sensitive,accurate and reproducible,and could provide an efficient way for the determination of the content of free glycerol in biodiesel.
Keywords:high performance liquid chromatography(HPLC)  evaporative light-scattering detector(ELSD)  biodiesel  glycerol  