首页 | 官方网站   微博 | 高级检索  
     


3D true-phase polarity recovery with independent phase estimation using three-tier stacks based region growing (3D-TRIPS)
Authors:Haining Liu  Gregory J Wilson  Niranjan Balu  Jeffrey H Maki  Daniel S Hippe  Wei Wu  Hiroko Watase  Jinnan Wang  Martin L Gunn  Chun Yuan
Affiliation:1.Department of Bioengineering,University of Washington,Seattle,USA;2.Department of Radiology,University of Washington,Seattle,USA;3.Department of Radiology,University of Colorado,Aurora,USA;4.Department of Radiology, Tongji Hospital,Huazhong University of Science and Technology,Wuhan,China;5.Department of Surgery,University of Washington,Seattle,USA;6.Department of Bioengineering,Tsinghua University,Beijing,China
Abstract:

Objectives

A postprocessing technique termed 3D true-phase polarity recovery with independent phase estimation using three-tier stacks based region growing (3D-TRIPS) was developed, which directly reconstructs phase-sensitive inversion-recovery images without acquisition of phase-reference images. The utility of this technique is demonstrated in myocardial late gadolinium enhancement (LGE) imaging.

Materials and methods

A data structure with three tiers of stacks was used for 3D-TRIPS to directly achieve reliable region growing for successful background-phase estimation. Fifteen patients undergoing postgadolinium 3D phase-sensitive inversion recovery (PSIR) cardiac LGE magnetic resonance imaging (MRI) were recruited, and 3D-TRIPS LGE reconstructions were compared with standard PSIR. Objective voxel-by-voxel comparison was performed. Additionally, blinded review by two radiologists compared scar visibility, clinical acceptability, voxel polarity error, or groups and blurring.

Results

3D-TRIPS efficiently reconstructed postcontrast phase-sensitive myocardial LGE images. Objective analysis showed an average 95% voxel-by-voxel agreement between 3D-TRIPS and PSIR images. Blinded radiologist review demonstrated similar image quality between 3D-TRIPS and PSIR reconstruction.

Conclusion

3D-TRIPS provided similar image quality to PSIR for phase-sensitive myocardial LGE MRI reconstruction. 3D-TRIPS does not require acquisition of a reference image and can therefore be used to accelerate phase-sensitive LGE imaging.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23