微小假体磨损颗粒诱导置入物旁骨溶解的扫描电镜观察
引用本文:王友.微小假体磨损颗粒诱导置入物旁骨溶解的扫描电镜观察[J].中华骨科杂志,1999,19(10):617-619.
作者姓名:王友
作者单位:上海第二医科大学附属第九人民医院骨科
摘    要:为比较不同人工关节微小磨损颗粒对假体-骨界面处骨整合及骨整合及骨溶解的影响,探讨假体松动的机制。方法本实验采用扫描电镜技术对钛事金、钴-铬-钼与聚乙烯(ultrahighmolecularweightpolyethylene,UHMWPE)三种微小颗粒诱导的假体周围骨组织改变进行超微结构观察。结果研究发现,直径2.5μm的Ti-6Al-4V颗粒诱导置入物旁骨吸收或骨溶解的程度明显低于相同直径的Co

关 键 词:假体植入  骨溶解  骨整合  人工关节  扫描电镜

Ultrastructure Observation on Periimplant Osteolysis Induced by Micron-sized Wear Particles
Abstract:Objective To compare the effects of various kinds of micron-sized wear particles on osteointegration at implant-bone interface and process of periimplant osteolysis, to delineate the mechanism of aseptic loosening of artificial joint. Methods Three special particles of ultra high molecular weight polyethylene(UHMWPE), Ti 6Al 4V and Co Cr Mo were used respectively to induce the change of periimplant bone structure of a well fixed implant made of titanium alloy in rabbit femur or tibia. The bone structure at implant-bone interface was observed by scanning electron microscope(SEM). Results Ti 6Al 4V particle induced less periimplant bone surface resorption or osteolysis than that by UHMWPE or Co Cr Mo particles. Conclusion Various kinds of wear particles with the same size and same amount show different capability in inducing periimplant osteolysis. Osteointegration seems to be more vulnerable to UHMWPE or Co Cr Mo particles.
Keywords:Prosthesis  Osseointegration  Microscopy  electron  scanning