首页 | 官方网站   微博 | 高级检索  
     


Method for bias field correction of brain T1-weighted magnetic resonance images minimizing segmentation error
Authors:Gispert Juan D  Reig Santiago  Pascau Javier  Vaquero Juan J  García-Barreno Pedro  Desco Manuel
Affiliation:Laboratorio de Imagen Médica, Medicina y Cirugía Experimental, Hospital General Universitario Gregorio Mara?ón, Madrid, Spain.
Abstract:This work presents a new algorithm (nonuniform intensity correction; NIC) for correction of intensity inhomogeneities in T1-weighted magnetic resonance (MR) images. The bias field and a bias-free image are obtained through an iterative process that uses brain tissue segmentation. The algorithm was validated by means of realistic phantom images and a set of 24 real images. The first evaluation phase was based on a public domain phantom dataset, used previously to assess bias field correction algorithms. NIC performed similar to previously described methods in removing the bias field from phantom images, without introduction of degradation in the absence of intensity inhomogeneity. The real image dataset was used to compare the performance of this new algorithm to that of other widely used methods (N3, SPM'99, and SPM2). This dataset included both low and high bias field images from two different MR scanners of low (0.5 T) and medium (1.5 T) static fields. Using standard quality criteria for determining the goodness of the different methods, NIC achieved the best results, correcting the images of the real MR dataset, enabling its systematic use in images from both low and medium static field MR scanners. A limitation of our method is that it might fail if the bias field is so high that the initial histogram does not show bimodal distribution for white and gray matter.
Keywords:nonuniform intensity correction  NIC  magnetic resonance imaging  bias field  intensity inhomogeneities  segmentation algorithms
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号