首页 | 官方网站   微博 | 高级检索  
     


Chromium(III) ion-selective electrode based on 4-dimethylaminoazobenzene
Authors:Abbaspour A  Izadyar A
Affiliation:Chemistry Department, Shiraz University, Shiraz, Iran. abbaspour@chem.susc.ac.ir
Abstract:A PVC-based membrane of 4-dimethylaminoazobenzene reveals a Nernstian potentiometric response (with slope of 19.5+/-0.6 mV/decade and a correlation coefficient of 0.999) for Cr(III) over a wide concentration range (1.66 x 10(-6)-1.0 x10(-2) mol dm(-3)). The potential of this electrode is independent of pH in the range of 3.0-5.5. It has a fast response time of about 10 s and was used for a period of 3 months with good reproducibility. The detection limits of this membrane electrode was 8 x 10(-7) M. the proposed electrode has been used as an indicator electrode in the potentiometric titration of Cr(III) with EDTA. This sensor exhibits a very good selectivities for Cr(III) over a wide variety of metal ions.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23