Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

2.
Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

3.
Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

4.
Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

5.
Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

6.
Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

7.
Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

8.
Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

9.
Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

10.
Objective To evaluate the clinical application of automated urine formed elements analyzer and/or urine dipstick analyzer for examination of urinary formed elements in screening urinary tract infection (UTI). Methods 148 fresh midstream clear-catch urine samples from the UTI patients and 284 fresh midstream clear-catch urine samples from non-UTI subjects were selected. Bacteria culture was performed for bacterial colony counting and identification. Bacteria counts ( BACT), yeast-like fungus and WBC were performed by UF-looOi automated urine formed elements analyzer. Leukocyte esterase test (LEU) and nitrite test (NIT) were performed by URISYS 2400 urine dipstick analyzer. We evaluated data obtained from urine dipstick analyzer, UF-1000i and combination of UF-1000i with urine dipstick analyzer and the results was compared with those obtained from quantitative bacterial culture. Then we evaluated the sensibility, specificity, positive predictive value (PPV), negative predictive value (NPV) and accuracy. Results Among the 148 patients with UTI, the positive rate of the quantitative bacterial culture was 73.6% (109/148), the positive rate of LEU and NIT detected by dipstick test 26. 4% (39/148).There was significantly statistical difference between bacterial culture and strip test(χ2 = 55.68 ,P < 0. 05 ). The positive rate of urine flow cytometry by UF-1000i with either positive of BACT and WBC was 91.2%(135/148), which was higher than the positive rate of the quantitative bacterial culture. There was significant difference between two methods (χ2 = 14. 70, P < 0. 05 ). The positive rate of anyone positive among BACT, WBC, LEU and NIT was 94. 6% (140/148) when detected with combination of dipstick test and UF-1000i, which was higher than the positive rate of the quantitative bacterial culture. And there was significant difference between two methods (χ2 = 20. 45, P < 0. 05 ). The sensitivity of dipstick test was low (26. 4% ,39/148 ), and specificity was high ( 99. 3%, 282/284 ) . The sensitivity, specificity, positive predictive value, negative predictive value of BACT detected by UF-1000i in diagnosing urinary tract infection were 92. 6% ( 137/148 ), 39. 8% ( 113/284 ). 44. 5% ( 137/308 ) and 91.1% ( 113/124 ), respectively. If the dipstick test was combined with UF-1000i, the sensitivity, negative predictive value, specificity, positive predictive value and accuracy were 98.0% ( 145/148 ), 97.1% ( 100/103 ). 35.2% (100/284) ,44. 1% (145/329) and 56. 7% (245/432), respectively. Conclusions The combination of urine dipstick test and automated urine formed elements analyzer UF-1000i plays an important role in early diagnosis of UTI. And it has significant value in diagnosis of UTI, especially for the patients with negative bacterial cultures of urine sample.  相似文献   

11.
目的 制定针对UF-1000i尿液分析流水线(由UF-1000i尿流式分析仪和AX-4030尿干化学分析仪组成)自动化检测结果的尿液镜检复检规则.方法 收集2009年9月至2010年2月解放军总医院尿液常规标本共2 839份.首先利用UF-1000i尿液分析流水线对2 839份尿液标本进行有形成分分析(包括RBC、WB...  相似文献   

12.
目的 分析尿有形成分分析仪、尿干化学分析仪及显微镜尿沉渣结果 的差异,建立显微镜复检规则.方法 以UF-1000i尿有形成分分析仪及AX-4280尿干化学分析仪构成检测流水线,对临床尿标本先用流水线进行红细胞(RBC)、白细胞(WBC)及管型(CAST)检测,再用显微镜进行结果 验证.结果 UF-1000i检测RBC、WBC灵敏度高于AX-4280,但特异度低于AX-4280;流水线检测尿标本的阴性预测值达100.00%.UF-1000i分析仪与显微镜尿沉渣WBC、RBC、CAST检测结果 存在明显差异(χ2值分别为16.41、13.47、16.41,P<0.05).建立适合该流水线的显微镜复检规则.结论 UF-1000i和AX-4280分析仪检测结果 有可能存在差异,此时不能以UF-1000i代替显微镜尿沉渣检查,需参照相应的复检规则进行复检,保证检测结果 准确性.  相似文献   

13.
王延群  公衍文 《检验医学》2011,26(12):858-860
目的探讨Sysmex UF-1000i全自动尿液有形成分分析仪(简称UF-1000i)对尿液中红细胞(RBC)、白细胞(WBC)和管型(CAST)检测的敏感性。方法对1 054例患者的尿液分别用UF-1000i、URISys-2400全自动干化学分析仪(干化学法)及Diasys R/S2005定量分析工作站(镜检法)3种方法进行分析,并比较3种方法对RBC、WBC、CAST的检测敏感性。结果 UF-1000i检测RBC、WBC、CAST的阳性率分别为20.6%、20.3%、6.7%;干化学法检测出RBC、WBC的阳性率分别为24.5%、17.9%;镜检法检出RBC、WBC、CAST的阳性率分别为16.9%、20.5%、2.2%。3种方法RBC、CAST的检出率差异有统计学意义(P〈0.01),WBC的检出率差异无统计学意义(P〉0.05)。结论 UF-1000i不能完全取代尿沉渣镜检,建议3种方法联合应用以减少检验误差,提高尿液分析质量。  相似文献   

14.
目的调查天津地区UF-50尿沉渣分析仪的正常参考值。方法用尿常规分析仪和尿液离心镜检两种方法筛选出正常尿液标本704份,用UF-50尿沉渣分析仪测定白细胞、红细胞、上皮细胞、管型、细菌的值,用百分位数法确定参考值。结果在95%可信区间的UF-50尿沉渣分析仪的参考范围为:红细胞0~14.8个/微升,白细胞0~17.3个/微升,上皮细胞0~14.7个/微升,管型0~0.44个/微升,细菌0~2642.3个/微升。结论确定的参考范围基本可信,应用UF-50进行尿沉渣检查结果可靠,可广泛应用于临床泌尿系统疾病的辅助诊断。  相似文献   

15.
目的探讨实验室内不同全自动尿沉渣分析仪检测结果之间的差异是否在允许范围之内,以提高检验结果的准确性、一致性、可比性。方法以Sysmex UF-1000i全自动尿沉渣分析仪(LJ-25)作为标准仪器,在两台全自动尿沉渣分析仪均处于正常状态下,双份检测新鲜尿液,获取红细胞计数(RBC/μL)、白细胞计数(WBC/μL)、上皮细胞计数(EC/μL)、管型计数(CAST/μL)及细菌计数(BACT/μL),将检测结果进行通过率计算及比对试验,并参照Sysmex公司的仪器说明书提供的标准作为各参数的允许范围进行评估。结果两仪器检测结果显示,RBC、WBC、EC、CAST及BACT比对通过率88.09%、88.09%、92.86%、52.38%、35.71%;线性相关系数(r)为0.998、0.997、0.993、0.417、0.971,CAST、BACT结果比较差异无统计学意义(P〉0.05),RBC、WBC、EC结果比较差异有统计学意义(P〈0.05)。结论临床实验室应加强对所有检测项目结果的可靠性进行验证,做好尿沉渣分析仪室内质控的同时定期进行仪器比对试验,以保证不同仪器上的检测结果具有可比性、准确性、一致性、持续性的质量保证。  相似文献   

16.

Objectives

The purpose of this Italian multicenter study was to define pediatric upper reference values for urine particle quantification by using automated flow cytometry.

Design and methods

Four hospital-based clinical laboratories participated in this multicenter investigation, which included a total study population of 161 Italian children aged from 1 to 12 years. Two laboratories used Sysmex UF-100 and analyzed 86 children, whereas the other two used Sysmex UF-1000i and analyzed 75 subjects. Particle quantification included the analysis of white blood cells (WBC), red blood cells (RBC), squamous epithelial cells (EC), transitional epithelial cells (TC), casts (CAST) and bacteria (BACT).

Results

The upper reference values in subjects tested with the Sysmex UF-100 were 9.7 WBC/μL, 10.1 RBC/μL, 7.5 EC/μL, 2.5 TC/μL, 0.7 CAST/μL and 3090 BACT/μL, whereas the upper reference values in subjects tested with the Sysmex UF-1000i were 10.5 WBC/μL, 8.3 RBC/μL, 7.2 EC/μL, 2.9 TC/μL, 0.7 CAST/μL and 48 BACT/μL. No statistically significant differences between genders were found in the value distribution of any of the parameters tested. Similarly, no statistically significant differences were observed between the two urine analyzers, except for BACT.

Conclusions

Automated analysis of urine particles appears a suitable means to optimize the workflow of routine urinalysis of children specimens. The upper reference limits for pediatric subjects obtained in this study were comparable to those previously reported in the literature, with no significant differences between genders and analyzers.  相似文献   

17.
目的探讨泰安地区健康成人尿液中白细胞、红细胞、上皮细胞的定量参考范围。方法选择1860例健康查体者,使用一次性定量计数板和全自动尿沉渣分析仪uf-100,直接计数随机尿中红细胞、白细胞和上皮细胞数量。结果定量计数板法男性:红细胞0~5.5/ul,白细胞0~9/ul,上皮细胞0~4/ul。女性:红细胞0~9/ul白细胞0~22/ul上皮细胞0~32/ul。UF-100法男性:红细胞0~8/ul白细胞0~12/ul上皮细胞0~8/ul女性红细胞0~12/ul自细胞0~24/ul上皮细胞0~32.5/ul两法参考范围接近。结论为泰安地区尿液有形成份检测规范化标准化和临床应用提供依据。  相似文献   

18.
目的对比分析AVE-763B和UF-1000i两种全自动尿沉渣分析仪检测尿液红细胞(RBC)、白细胞(WBC)、管型、结晶结果,并与标准人工镜检进行比较,评价两种仪器的检测性能的差异。方法随机选取门诊各科室送检的新鲜中段尿标本150例,分别用AVE-763B和UF-1000i全自动尿沉渣分析仪以及人工镜检对WBC、RBC、管型、结晶进行计数,并以显微镜镜检为标准,对检测结果进行比对分析。结果 AVE-763B、UF1000i两种尿沉渣分析仪检测尿液中RBC与人工镜检符合率为91.33%、95.33%,灵敏度为80.95%、97.62%,特异性为95.37%、94.44%;WBC与人工镜检符合率为:89.33%、91.30%,灵敏度为92.31、90.77%,特异性为87.06%、91.76%;管型与人工镜检符合率为95.33%、92.67%,灵敏度为75.00%、87.50%,特异性为96.48%、92.96.00%;结晶与人工镜检符合率为96.00%、98.00%,灵敏度为83.33%、83.33%,特异性为96.52%、98.61%。AVE-763B和UF-1000i尿沉渣分析仪检测RBC携带污染率为0.03%和0.75%、WBC携带污染率为0.08%和0.16%。AVE-763B和UF-1000i两种全自动尿沉渣分析仪检测结果的差异无统计学意义(P>0.05)。结论 AVE-763与UF-1000i全自动尿沉渣分析仪灵敏度较高,携带污染率低,可用于标本初筛,但仍需要配合人工镜检以提高检验质量。  相似文献   

19.
三种方法在尿液有形成分分析中的对比观察   总被引:2,自引:0,他引:2  
目的 探讨UF-1000i尿液有形成分分析仪与干化学方法、手工显微镜检查尿中红细胞(RBC)、白细胞(WBC)的符合情况及影响因素.方法 使用UF-1000i 尿有形成分分析仪、干化学分析仪与显微镜分别检测996例尿液标本中RBC、WBC,将3种方法的检测结果进行比较.结果 分别以RBC、WBC超过25和30个/μL为...  相似文献   

20.
目的评价UF-1000i尿流式有形成分分析仪在尿路感染(UTI)筛查中的应用价值。方法应用UF—1000i尿流式有形成分分析仪检测148例疑似uTI患者中段尿标本中的细菌(BACT)数和白细胞(WBC)数,以定量尿细菌培养作为诊断UTI金标准,应用受试者工作特征(ROC)曲线确立阳性判断标准,评价UF-1000i尿流式有形成分分析仪诊断UTI的敏感性、特异性、阳性预测值、阴性预测值和准确性。结果UF—1000i尿有形成分分析仪BACT和WBC诊断UTI的cut—off值分别为325个/汕和48个/μL,其诊断UTI的敏感性、特异性、阳性预测值、阴性预测值、准确性分别是:78.3%、92.2%、81.8%、90.4%、87.8%和73.9%、81.4%、64.2%、87.4%、79.1%。UF—1000i联合诊断uTI的敏感性、特异性、阳性预测值、阴性预测值和准确性分别95.7%、70.6%、59.5%、97.3%、78.4%。结论UF.1000i尿有形成分分析仪可作为UTI的快速筛查工具。  相似文献   