CT扫描参数对人体组织CT值影响的研究
引用本文:彭文献,彭天舟,叶小琴,付益谋,潘慧平,高源统,金光波.CT扫描参数对人体组织CT值影响的研究[J].中华放射医学与防护杂志,2010,30(1):79-81.
作者姓名:彭文献  彭天舟  叶小琴  付益谋  潘慧平  高源统  金光波
作者单位:浙江省瑞安人民医院(温州医学院附属第三医院)放射科,瑞安,325200
基金项目:温州市科技局科研基金 
摘    要:目的 探索不同CT扫描参数对人体同一种组织CT值的影响。方法 通过在同一台CT机上,分别改变其中1个扫描参数,如X线管电压、毫安秒和重建函数等,而保持其他扫描参数不变,多次扫描标准体模,测量和分析体模中不同物质的CT值。结果 X线管电压的改变对物质的CT值影响具有显著意义。聚乙烯、聚碳酸酯、有机玻璃的CT值与管电压成正相关;聚四氟乙烯的CT值与管电压成负相关。毫安秒和重建函数对CT值的影响差异无统计学意义。结论 同一个人体组织在不同的X线管电压条件下CT值是变化的。因此,在临床影像诊断和放疗中应该考虑图像扫描参数的设置对诊断和治疗结果的影响。

关 键 词:CT值  X线管电压  扫描参数  相关性
收稿时间:2009/6/28 0:00:00

Effect of CT scanning parameters on CT number
PENG Wen-xian,PENG Tian-zhou,YE Xiao-qin,FU Yi-mou,PAN Hui-ping,GAO Yuan-tong and JIN Guang-bo.Effect of CT scanning parameters on CT number[J].Chinese Journal of Radiological Medicine and Protection,2010,30(1):79-81.
Authors:PENG Wen-xian  PENG Tian-zhou  YE Xiao-qin  FU Yi-mou  PAN Hui-ping  GAO Yuan-tong and JIN Guang-bo
Affiliation:Rui'an People Hospital, Rui'an 325200, China;Rui'an People Hospital, Rui'an 325200, China;Rui'an People Hospital, Rui'an 325200, China;Rui'an People Hospital, Rui'an 325200, China;Rui'an People Hospital, Rui'an 325200, China;Rui'an People Hospital, Rui'an 325200, China;Rui'an People Hospital, Rui'an 325200, China
Abstract:Objective To study the effects on tissue CT number caused by scan protocols. Methods The phantom was repeatedly scanned in different protocols by changing only one of parameters, such as X-ray tube voltage,mAs and recon kernel, while other parameters were ketp unchanged. The CT number of different materials in phantom were measured and analyzed. Results The CT numbers of tissues changed remarkably with the tube voltage and had different relativity for different tissues. The CT numbers had positive correlation with kV for such maierials as polyethyle, lexan, perspex,but for teflon the correlation was negative. The mAs and recon kernel had no effects on CT number. Conclusions The CT number of tissue changes with scanning X-ray tube voltage, so the setting of scan parameters should be taken into account in image diagnosis and radiotherapy.
Keywords:CT number  X-Ray tube voltage  Scan parameter  Relativity