多通道剂量测量仪在调强放疗剂量检测中的应用
引用本文:李珠明,崔玉琴,白晓平,常青,张海芹,王新花.多通道剂量测量仪在调强放疗剂量检测中的应用[J].中国医学物理学杂志,2014(1):4600-4603.
作者姓名:李珠明  崔玉琴  白晓平  常青  张海芹  王新花
作者单位:包头肿瘤医院放疗科,内蒙古包头014030
基金项目:内蒙古自治区卫生厅医疗卫生科研计划项目(2010244)
摘    要:目的:保证调强放疗治疗计划临床实施的正确性。方法:分别用0coy、50cGy、100cGy、150cGy、200cGy、250cGy、300cGy照射多通道剂量仪金属氧化物半导体场效应晶体管(MOSFET)探头,观察多通道剂量仪读数;大机架分别旋转0、30、60、90、120、150、180、210、240、270、300度时,计划照射探头100coy时,观察多通道剂量仪读数;采用多通道剂量仪验证空间多点绝对剂量。结果:计划剂量与多通道剂量仪探头阈值电压之间呈正相关关系(r=l,P=-0.0000):计划剂量与多通道剂量仪探头读数剂量之间呈正相关关系,(r=1,P=0.0000);测得多通道剂量仪MOSFET探头,轴方向性误差〈2%:在模体中测量绝对剂量,等中心点实测剂量与计划剂量误差在3%以内,其它偏离点的实测剂量与计划剂量误差在5%以内。结论:应用多通道剂量仪测量IMRT绝对剂量是可行的.可有效地保证IMRT的顺利开展。

关 键 词:调强放疗  多通道剂量仪  绝对剂量验证

Absolute Dose Verification of Intensity Modulated Radiation Therapy with Multi-channel Dose Measure Iinstrument
LI Zhu-ming,CUI Yu-qin,BAI Xiao-ping,CHANG Qing,ZHANG Hai-qin,WANG Xin-hua.Absolute Dose Verification of Intensity Modulated Radiation Therapy with Multi-channel Dose Measure Iinstrument[J].Chinese Journal of Medical Physics,2014(1):4600-4603.
Authors:LI Zhu-ming  CUI Yu-qin  BAI Xiao-ping  CHANG Qing  ZHANG Hai-qin  WANG Xin-hua
Affiliation:(Department of radiation oncology, Baotou Tumor Hospital, Baotou Inner Mongola 014030, Chain)
Abstract:Objective: To ensure the accuracy of intensity modulated radiation therapy. Methods: The MOSFET probe of mul- ti-channel dose measure instrument was radiated by 0 cGy, 50 cGy, 100 cGy, 150 cGy, 200 cGy, 250 eGy and 300 cGy, abso- lute dose was measured with multi-channel dose measure instrument, the MOSFET was radiated by 100 cGy where the gantry was on 0, 30, 60, 90, 120, 150, 180, 210, 240, 270 and 300 angle, absolute dose was measured with the multi-channel dose mea- sure ins~ment, Results: The threshold voltage of MOSFET and the plan dose was positive correlation (r=l, P=-0.0000), the plan dose and indication dose of the multi-channel dose measure instrument was positive correlation (r=l, P=-0.0000), the toler- ance measured with the multi-channel dose measure instrument was less than 2% where the gantry on different angle. The errors between the indication dose and plan dose in the central point were less than 3%, whereas the errors in another eccentric point were less than 5%. Conclusions: Absolute dose of IMRT may be measured with the multi-channel dose measure instrument and ensure quality of IMRT.
Keywords:intensity modulated radiation therapy  multi-channel dose measure instrument  absolute dose verification