首页 | 官方网站   微博 | 高级检索  
     

离子色谱-双波长紫外检测法同时测定水中碘酸盐和碘离子的含量
引用本文:李 锦,秦晓娟,付志斌,刘玉欣,杨凤艳,燕 阔.离子色谱-双波长紫外检测法同时测定水中碘酸盐和碘离子的含量[J].食品安全质量检测技术,2021,12(5):1736-1740.
作者姓名:李 锦  秦晓娟  付志斌  刘玉欣  杨凤艳  燕 阔
作者单位:河北省疾病预防控制中心,河北省疾病预防控制中心,河北省疾病预防控制中心,河北省疾病预防控制中心,沧州市疾病预防控制中心,河北省疾病预防控制中心
摘    要:目的建立离子色谱-双波长紫外检测法同时测定水中碘酸盐和碘离子的分析方法。方法水样过0.22μm滤膜后直接进样测定,水中碘酸根和碘离子通过离子色谱柱分离,紫外检测器分别在200nm和226nm处检测。结果碘酸根和碘离子在0.01~1.00mg/L浓度范围内线性良好,相关系数r分别为0.99953和0.99950,检出限为0.004和0.005mg/L,相对标准偏差(relative standard deviations,RSDs)为2.8%和2.7%(n=8)。低(0.05mg/L)、中(0.10mg/L)、高(0.50mg/L)3个浓度IO_3~-和I~-加标回收率范围为94%~106%和92%~102%,精密度为1.04%~4.82%和0.75%~4.47%。结论本方法操作简单、准确度高、精密度好、线性范围宽,适用于水中碘酸盐和碘离子的测定。

关 键 词:离子色谱-双波长紫外检测法    碘酸盐  碘离子
收稿时间:2020/11/10 0:00:00
修稿时间:2021/2/25 0:00:00

Simultaneous determination of iodate and iodine ions in water by ion chromatography dual wavelength ultraviolet detection method
LI Jin,QIN Xiao-Juan,FU Zhi-Bin,LIU Yu-Xin,YANG Feng-Yan,YAN Kuo.Simultaneous determination of iodate and iodine ions in water by ion chromatography dual wavelength ultraviolet detection method[J].Food Safety and Quality Detection Technology,2021,12(5):1736-1740.
Authors:LI Jin  QIN Xiao-Juan  FU Zhi-Bin  LIU Yu-Xin  YANG Feng-Yan  YAN Kuo
Affiliation:Hebei Provincial Center for Disease Control and Prevention,Hebei Provincial Center for Disease Control and Prevention,Hebei Provincial Center for Disease Control and Prevention,Hebei Provincial Center for Disease Control and Prevention,Hebei Provincial Center for Disease Control and Prevention,Hebei Provincial Center for Disease Control and Prevention
Abstract:Objective To establish a method for simultaneous determination of iodate and iodine ions in water by ion chromatography-dual wavelength ultraviolet detection.Methods After filtered through 0.22μm filter,the water sample was directly injected for determination.The iodate and iodine ions in water were separated by ion chromatography column and detected by ultraviolet detectorat 200 nm and 226 nm.Results The iodate and iodide ions were linear in the concentration range of 0.01-1.00 mg/L.The correlation coefficients(r 2)were 0.99953 and 0.99950,limits of detection were 0.004 mg/L and 0.005 mg/L,relative standard deviations(RSDs)were 2.8%and 2.7%(n=8).The recoveries of IO3-and Iat low(0.05 mg/L),medium(0.10 mg/L)and high(0.50 mg/L)concentration levels were 94%?106%and 92%?102%,and the precisions were1.04%?4.82%and 0.75%?4.47%.Conclusion This method has the advantages of simple operation,high accuracy,good precision and wide linear range,which is suitable for the determination of iodate and iodide ions in water.
Keywords:ion  chromatograph  dual  wavelength UV  detection  water  iodate  iodine  ion  simultaneous  determination
本文献已被 维普 等数据库收录!
点击此处可从《食品安全质量检测技术》浏览原始摘要信息
点击此处可从《食品安全质量检测技术》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号