胸部X射线CT扫描中的放射防护最优化
引用本文:李士骏,朱晓华.胸部X射线CT扫描中的放射防护最优化[J].辐射防护,2006,26(3):129-136.
作者姓名:李士骏  朱晓华
作者单位:1. 苏州大学放射医学与公共卫生学院,215123
2. 上海市肺科医院影像科,200433
摘    要:X射线螺旋CT机不同扫描条件下,用有机玻璃均匀模体测量多层扫描的CT剂量指数值,用均匀水模体测量CT值的均匀性及其噪声。比较了不同扫描条件下,患者肺部肿块影及片状影的CT图像质量。结果表明:曝露量为40、25、7.5mAs时的CT剂量指数值分别不超过曝露量为115mAs时的40%、30%和15%。本文用CT扫描“优化因子”作为图像质量、病人所受剂量的综合评价指标,实现了CT扫描条件优化的定量分析。虽然,随X射线曝露量减少,均匀模体断层内CT值噪声增加、正常图像百分率下降,但从图像质量、病人剂量综合考虑,X射线胸部CT扫描的曝露量宜取25mAs。

关 键 词:X射线CT  胸部扫描  曝露量  CT剂量指数  CT值均匀性  CT值噪声  CT扫描优化因子
收稿时间:2005-03-14
修稿时间:2005-03-14

OPTIMIZATION OF RADIATION PROTECTION IN X-RAY CT SCAN OF CHEST
Li Shijun,Zhu Xiaohua.OPTIMIZATION OF RADIATION PROTECTION IN X-RAY CT SCAN OF CHEST[J].Radiation Protection,2006,26(3):129-136.
Authors:Li Shijun  Zhu Xiaohua
Affiliation:1. School Of Radiation Medicine and Public Health, Soochow University, Suzhou, 215123;2. Department of Radiology, Shanghai Pneumology Hospital, Shanghai, 200433
Abstract:Under different scan conditions,CT dose indices of multiple scan were measured using homogeneous PMMA phantom.Homogeneity and noise of CT values were also measured using a water phantom.CT image quality for mass and patch in the lung of patients that were selected randomly was compared.The results indicated: that CT dose indices for 40,25,7.5 mAs are not more than 40%,30%,15% of that for 115 mAs respectively.Quantitative analysis of scan conditions was carried out by using CT scan optimization factor as a comprehensive index to evaluate image quality and patient dose.The noise of CT values in homogeneous phantom increased and percentage of normal image decreased with reduction of X-ray exposure.Considering image quality and patient dose comprehensively,the optimized exposure for chest scanning is 25 mAs.
Keywords:X-my CT  Chest Scanning  Exposure  CT Dose Index  Homogeneity of CT Value  Noise of CT Value  CT Scan Optimization Factor