Characteristics of parallel-hole and pinhole collimators for nuclear medicine imaging
Authors:LIU Silu  ZHU Jie
Affiliation:[1]The Second Artillery Equipment Academy ofPLA, Beijing 100085, China [2]96261 Unit of Chinese People's Liberation Army, Lingbao 472500, China
Abstract:In this paper,the spatial resolution and geometry sensitivity of parallel- and pin-hole collimators in compact gamma camera are presented,and quantitatively compared by GEANT4 code in Monte-Carlo library.The results show that the geometry sensitivity for pinhole collimator rapidly drops with increasing the gamma source-to- collimator distance(SCD);and for parallel-hole collimator,mildly.Meanwhile,the spatial resolution for pin-hole collimator mildly deteriorates;and for parallel-hole collimator,severely.The pin-hole collimator for close imaging objects has higher geometry sensitivity than parallel-hole collimator.Our findings are helpful for setting compact gamma camera collimators in nuclear medicine imaging.
Keywords:Spatial resolution  Geometric sensitivity  Collimator  Nuclear medicine