首页 | 官方网站   微博 | 高级检索  
     


Monte Carlo simulation of breast tumor imaging properties withcompact, discrete gamma cameras
Authors:Gruber  GJ Moses  WW Derenzo  SE
Affiliation:Lawrence Berkeley Lab., CA;
Abstract:Describes Monte Carlo simulation results for breast tumor imaging using a compact, discrete gamma camera. The simulations were designed to analyze and optimize camera design, particularly collimator configuration and detector pixel size. Simulated planar images of 5-15 mm diameter tumors in a phantom patient (including a breast, torso, and heart) were generated for imaging distances of 5-55 mm, pixel sizes of 2×2-4×4 mm2, and hexagonal and square hole collimators with sensitivities from 4000 to 16,000 counts/mCi/sec. Other factors considered included T/B (tumor-to-background tissue uptake ratio) and detector energy resolution. Image properties were quantified by computing the observed tumor FWHM (full-width at half-maximum) and S/N (sum of detected tumor events divided by the statistical noise). Results suggest that hexagonal and square hole collimators perform comparably, that higher sensitivity collimators provide higher tumor S/N with little increase in the observed tumor FWHM, that smaller pixels only slightly improve tumor FWHM and S/N, and that improved detector energy resolution has little impact on either the observed tumor FWHM or the observed tumor S/N
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号