首页 | 官方网站   微博 | 高级检索  
     


Biocompatibility study of plasma-coated nitinol (NiTi alloy) stents
Authors:Wang G  Shen Y  Cao Y  Yu Q  Guidoin R
Affiliation:Chongqing University, Bioengineering College, Chongqing, People's Republic of China. guixue_wang@126.com
Abstract:The authors aimed to assess the surface modification effects of plasma coatings on biocompatibility of nitinol intravascular stent in terms of anticoagulation, haemocytolysis rate, hydrophilicity, cytotoxicity and so on. In order to improve their surface adhesive properties to endothelial cells, NiTi alloy intravascular stents were treated and coated using a low-temperature plasma deposition technique. It was found that plasma coating changed the surface morphology of the stents to a micron-level surface roughness in the range of 1-5 microm. In comparison with the untreated control, the plasma-treated NiTi alloy intravascular stents showed increased surface hydrophilicity and enhanced anticoagulation property. Testing results on plasma-coated NiTi stents indicated that they complied with the standard of national biologic safety evaluation of medical apparatus and instrument (GB/T16886-1997, People's Republic of China) in terms of haemocytolysis rate, cytotoxicity and pyretogen.
Keywords:
本文献已被 PubMed 等数据库收录!