APPLICATIONS OF CFD TECHNIQUE IN THE DESIGN AND FLOW ANALYSIS OF IMPLANTABLE AXIAL FLOW BLOOD PUMP
Authors:FAN Hui-min  HONG Fang-wen  ZHANG Guo-ping  YE Liang  LIU Zhong-min
Affiliation:[1]Shanghai East Hospital Affiliated to Tongji University, Shanghai 200120, China [2]China Ship Scientific Research Center, Wuxi 214082, China
Abstract:Based on established numerical methods and hydrodynamic performance testing facilities, CFD technique are applied to improve the design of the implantable axial flow blood pump and the flow analysis. (1) Applying brushless machine magnet steel, reducing its thickness while increasing the length, the flow channel's cross-section is increased, with no space connection between the large and the small rotators, and with a cone transition segment from the bearing to the principal axis, the flow is made smoother. The rotating speed is lowered by 1000 rpm – 1200 rpm under the same flow rate and pressure head, and thus the hemolysis can be avoided. (2) Different outlet stator guiding vanes are selected for the same blood pump for analyses of hydrodynamic performances and flow fields. An excellent design not only can regulate the rotating flow field into an axial one, reduce the circulation loss and improve the pump efficiency, but also can avoid backflow, vortex and secondary flow at the pump outlet, and thus the thrombus can be prevented. (3) The calf live tests show that some residual clots exist at the inner wall of the outlet connection bends, which are analyzed and explained by CFD techniques and the corresponding improvements are proposed. All results are verified by hydrodynamic performance tests and PIV flow field tests, and consistent conclusions are obtained.
Keywords:axial flow blood pump  CFD  PIV
本文献已被 CNKI 维普 万方数据 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号