A novel approach for realization of primary vibration calibration standard by homodyne laser interferometer in frequency range of 0.1 Hz to 20 kHz
Authors:Naveen Garg  Omkar Sharma  Ashok Kumar  Mark I Schiefer
Affiliation:1. Acoustics, Ultrasonic and Vibration Standards, Apex Level Standards and Industrial Metrology Division, National Physical Laboratory, CSIR, New Delhi 110 012, India;2. The Modal Shop, Inc., 3149 East Kemper Rd., Cincinnati, OH 45241, USA
Abstract:The paper discusses the primary vibration calibration standard of NPL, India capable of calibrating the reference accelerometers in frequency range from 0.1 Hz to 20 kHz as per ISO 16063-11. The excitation subsystem produces constant vibration at a specified amplitude and frequency, while the measurement system uses NI interface for measuring the quadrature output. The acceleration level and voltage level at the calibration frequency f is determined by applying a Discrete Fourier Transform to the voltage and displacement signals, and then examining the spectral component at frequency f. A PC-based data acquisition system acquires the accelerometer voltage signal and analog quadrature interferometer photodetector signal pair as well as a digital quadrature pair whereby the software processes the demodulated photodetector signals to reconstruct the armature displacement. The validation of the calibration results for standard reference accelerometers with manufacturer results and uncertainty in calibration in entire frequency range 0.1 Hz to 20 kHz is reported in the present work.
Keywords:ISO 16063-11  Sine-approximation method  Discrete Fourier Transform (DFT)
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号