Validation of Platelet Counting Accuracy With the Celltac F Automated Hematology Analyzer
Authors:Yutaka Nagai  Hiroshi Kondo  Noriyuki Tatsumi
Affiliation:1. Nihon Kohden Corporation 1-31-4, Shinjuku, Tokyo, 161-8560, Japan.;2. Daito Bunka University, Saitama, 335-8501, Japan.;3. International Buddhist University, Osaka, 583-8501, Japan,
Abstract:Rapid and accurate analysis of platelet count plays an important role in evaluating hemorrhagic status. Therefore, we evaluated platelet counting performance of a hematology analyzer, Celltac F (MEK-8222, Nihon Kohden Corporation, Tokyo, Japan), that features easy use with low reagent consumption and high throughput while occupying minimal space in the clinical laboratory. All blood samples were anticoagulated with dipotassium ethylenediaminetetraacetic acid (EDTA-2K). The samples were stored at room temperature (18^C–22^C) and tested within 4 hours of phlebotomy. We evaluated the counting ability of the Celltac F hematology analyzer by comparing it with the platelet counts obtained by the flow cytometry method that ISLH and ICSH recommended, and also the manual visual method by Unopette (Becton Dickinson Vacutainer Systems). The ICSH/ISLH reference method is based on the fact that platelets can be stained with monoclonal antibodies to CD41 and/or CD61. The dilution ratio was optimized after the precision, coincidence events, and debris counts were confirmed by the reference method. Good correlation of platelet count between the Celltac F and the ICSH/ISLH reference method (r = 0.99, and the manual visual method (r= 0.93) were obtained. The regressions were y = 0.90 x+9.0 and y=1.11x+8.4, respectively. We conclude that the Celltac F hematology analyzer for platelet counting was well suited to the ICSH/ISLH reference method for rapidness and reliability.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号