Comparative Evaluation of the Implant Made of Yttrium-partially Stabilized Zirconia and Pure Titanium on Stress Distribution by Anisotropic Finite Element Method
作者姓名:满毅  宫苹
作者单位:West China Stomatology College Sichuan University Chengdu 610041 China,West China Stomatology College Sichuan University Chengdu 610041 China
基金项目:FundedbyAppliedandBasicResearchfromScienceandTechnologyBureauofSichuanProvince(No.D301SY05169)
摘    要:1Introduction Implantplacementindentistry1]andorthopedics2]isapredicabletreatmentmodality.Thematerialsforden talimplantsandjointsaremetalandpolyethylene.These materialshavebeenusedfordecades;however,thein creasingproblemssuchasthewearofthepolyethylene,t…


Comparative Evaluation of the Implant Made of Yttrium-partially Stabilized Zirconia and Pure Titanium on Stress Distribution by Anisotropic Finite Element Method
Abstract:The purpose of this research is to analyze and compare stress distribution patterns around dental implant made of pure titanium and yttrium-partial stabilized zirconia (YPSZ) in anisotropic versus isotropic finite element method under vertical and oblique loads. Although the properties of implant and crown were changed, similar stress distribution and close stress level were observed in two different implant finite element models. The stress values were a little lower in the YPSZ model. In the bone, anisotropy increased the stress values by 30%-70% in the isotropic cases. The YPSZ implant could be more valuable choice for implant because of esthetic requirement. Anisotropy had subtle, yet significant effects on the stress level.
Keywords:finite element  anisotropic elasticity  zirconia  implant