首页 | 官方网站   微博 | 高级检索  
     


Preparation of polymer-coated, scintillating ion-exchange resins for monitoring of 99Tc in groundwater
Authors:Seliman Ayman F  Samadi Azadeh  Husson Scott M  Borai Emad H  DeVol Timothy A
Affiliation:Department of Analytical Chemistry and Environmental Control, Atomic Energy Authority, Cairo, Egypt.
Abstract:The present study was oriented to prepare new scintillating anion-exchange resins for measurement of (99)TcO(4)(-) in natural waters. The organic fluor 2-(1-naphthyl)-5-phenyloxazole was diffused into (chloromethyl)polystyrene resin. Thereafter, a thin layer of poly2-(methacryloyloxy)ethyl]trimethylammonium chloride] was grafted from the resin surface by surface-initiated atom transfer radical polymerization as an attempt to overcome potential problems related to the leaching of fluor molecules during usage. The residual chloromethyl groups of the polymer-coated resin were aminated by reaction with two different tertiary amines, triethylamine (TEA) and methyldioctylamine (MDOA). Off- and on-line quantification of (99)Tc was achieved with high detection efficiencies of 60.72 ± 1.93% and 72.83 ± 0.81% for resin with TEA and MDOA functional groups, respectively. The detection limit was determined to be less than the maximum contaminant level (33 Bq L(-1)) established under the Safe Drinking Water Act. The two functionalized resins were demonstrated to be selective for pertechnetate from synthetic groundwater containing up to 1000 ppm Cl(-), SO(4)(2-), and HCO(3)(-) and up to 1200 ppb Cr(2)O(7)(2-) in an acidic medium.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23