最佳kVp与体模表面剂量和体模厚度关系的实验研究
引用本文:赫英辉,韩月东,高凯,邓振生,余厚军,牟轩沁,任东青,孙立军.最佳kVp与体模表面剂量和体模厚度关系的实验研究[J].中华放射医学与防护杂志,2006,26(3):275-277.
作者姓名:赫英辉  韩月东  高凯  邓振生  余厚军  牟轩沁  任东青  孙立军
作者单位:1. 710032,西安,第四军医大学西京医院放射科
2. 西安交通大学图像处理与模式识别研究所
3. 第四军医大学放射医学教研室
基金项目:国家“十五863”基金资助项目(2001AA114152);国家自然科学基金项目(60472004)
摘    要:目的在保持一定的X射线图像信噪比的前提下,探讨管电压(kVp)与体模表面剂量和体模厚度的关系.方法应用自制人体模型模拟X射线数字减影血管造影,在特定的X射线影像增强器表面剂量下,采用不同的kVp和毫安秒组合曝光,采集相关数据并绘制成体模表面剂量-kVp曲线、体模厚度-kVp曲线以及体模厚度和kVp的关系式.结果体模表面剂量-kVp曲线提示,kVp越大,体模表面剂量越低,但斜率渐趋于零;体模厚度-kVp曲线提示,体模越厚,所需kVp越大;体模厚度和kVp的关系式为:kVp=64+0.89D(D为体模厚度,cm).结论随着kVp的不断增高会导致体模表面剂量的逐渐降低,理想的kVp值与体模的厚度直接相关.

关 键 词:人体模型  表面剂量  厚度  信噪比  管电压
收稿时间:2005-09-02
修稿时间:2005年9月2日

The experimental study of the optimized kVp and it's relationships with the surface dose of phantom and the phantom thickness
HE Ying_hui,HAN Yue_dong,GAO Kai.The experimental study of the optimized kVp and it''s relationships with the surface dose of phantom and the phantom thickness[J].Chinese Journal of Radiological Medicine and Protection,2006,26(3):275-277.
Authors:HE Ying_hui  HAN Yue_dong  GAO Kai
Affiliation:Department of Radiology, Xijinghospital, Xi-an 710032, China
Abstract:Objective To investigate the relationships of X-ray tube voltage (kVp) with surface dose of the phantom and phantom-thickness on the basis of the particular signal to noise ratio of X-ray images. Methods X-ray digital subtraction angiography was simulated using a phantom made of acrylic acids. Under the particular surface dose of X-ray image intensifier, the curve of the surface doses of phantom surface-kVp and the curve of the phantom thickness-kVp were acquired with different exposure combinations of kVp and mAs. Results According to the curve of the surface dose of phantom-kVp, it suggestsd that the gradual increase in kVp corresponded to the gradually decreased surface dose of phantom, with the slope tending to zero. However, the gradual increased phantom thickness corresponded to the gradually increased kVp on the basis of the curve of the phantom thickness-kVp, with the presentation of kVp=64+0.89D (phantom thickness, cm) between the phantom thickness and kVp. Conclusion The graduall increased in kVp might result in the gradual decrease in surface dose of phantom, the optimized kVp relating directly to the phantom thickness.;
Keywords:Phantom  Surface dose  Thickness  Signal to noise ratio  X-ray tube voltage (kVp)