首页 | 官方网站   微博 | 高级检索  
     


Non-invasive headspace measurement for characterizing oxygen-scavenging in polymers
Authors:Hua Li  Benny D Freeman  Mark E Stewart  Thomas R Clark
Affiliation:a The University of Texas at Austin, Department of Chemical Engineering, Center for Energy and Environmental Resources, 10100 Burnet Road, Building 133, Austin, TX 78758, United States
b Global PET Technology, Eastman Chemical Company, P.O. Box 511, Kingsport, TN 37662, United States
c Packaging Strategic Research, Kraft Foods, 200 DeForest Avenue, East Hanover, NJ 07936, United States
Abstract:A non-invasive headspace analysis method to measure rate and amount of oxygen uptake in oxygen-scavenging polymers is presented. Oxygen uptake data for metal-catalyzed poly(1,4-butadiene) at 30 °C are provided to illustrate the method. These data were obtained by measuring oxygen headspace concentration above the scavenging polymer with an OxySense® 200T non-invasive oxygen sensor, and, for comparison, oxygen uptake was measured with an analytical balance. Excellent agreement was observed between these two independent experiments.
Keywords:Oxygen scavenging  Polybutadiene  Oxidation
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23