纤维与电子软性输尿管镜在泌尿外科的应用对比研究
引用本文:张敏,李芳芳,邓碧仪,冯秋霞,孙红玲.纤维与电子软性输尿管镜在泌尿外科的应用对比研究[J].国际医药卫生导报,2022,28(2):244-247.
作者姓名:张敏  李芳芳  邓碧仪  冯秋霞  孙红玲
作者单位:广州医科大学附属第一医院微创外科中心手术室,广州 510230
摘    要:目的:对电子软性输尿管镜与纤维软性输尿管镜在泌尿外科手术中的应用情况进行对比分析,以便其发挥最优效能。方法:对广州医科大学附属第一医院微创外科中心手术室在2011年1月至2018年12月间购入的7条奥林巴斯软性输尿管镜的故障维修情况进行回顾性分析。同时,以同一时间、同一组医生开始使用的电子软性输尿管镜(URF-V1,S...

关 键 词:电子软性输尿管镜  纤维软性输尿管镜  泌尿外科
收稿时间:2021-09-09

Comparative study on the application of fiber and electronic flexible ureteroscopes in urological surgeries
Zhang Min,Li Fangfang,Deng Biyi,Feng Qiuxia,Sun Hongling.Comparative study on the application of fiber and electronic flexible ureteroscopes in urological surgeries[J].International Medicine & Health Guidance News,2022,28(2):244-247.
Authors:Zhang Min  Li Fangfang  Deng Biyi  Feng Qiuxia  Sun Hongling
Affiliation:Operating Room, Minimally Invasive Surgery Center, The First Affiliated Hospital of Guangzhou Medical University, Guangzhou 510230, China
Abstract:Objective To compare and analyze the application of electronic flexible ureteroscope and fiber flexible ureteroscope in urological surgeries, in order to maximize their performance. Methods A retrospective analysis was conducted on the maintenance of 7 Olympus flexible ureteroscopes purchased in the Operating Room, Minimally Invasive Surgery Center, The First Affiliated Hospital of Guangzhou Medical University from January 2011 to December 2018. At the same time, the efficacies of electronic flexible ureteroscope (URF-V1, SN: 2240126) and fiber flexible ureteroscope (URF-P5, SN: 2344111), which were used by the same doctors at the same time, were retrospectively analyzed in terms of the causes, locations, and times of failures. URF-V1 (SN: 2240126) was used for 407 times, including 277 male cases and 130 female cases, with an average age of 48.93 years old; URF-P5 (SN:2344111) was used for 556 times, including 333 male cases and 223 female cases, with an average age of 49.21 years old. The cost of fiber and electronic flexible ureteroscopes in surgeries was calculated. Independent sample t test was used for the measurement data, and χ2 test was used for the count data. Results The flexible ureteroscopy was successfully performed in urological surgeries using URF-V1 (SN: 2240126) or URF-P5 (SN: 2344111). The intraoperative performance of lens was stable, the imaging effect was clear, and the doctors were satisfied with the effect. In the analysis of the failure maintenance report, it was found that water leakage of flexible ureteroscope was the most common failure reason, accounting for 56.5% (26/46) of all failure reasons, and water leakage of insertion channel accounted for 69.2% (18/26). In terms of surgery cost, URF-V1 (SN: 2240126) failed for 8 times, 1 times of maintenance every 51 times of usage, and the cost of each surgery was about 5 020 Yuan; URF-P5 (SN: 2344111) failed for 15 times, 1 times of maintenance every 37 times of usage, and the cost of each surgery was about 2 190 Yuan. Conclusions Both electronic and fiber flexible ureteroscopes are safe and effective methods for urinary surgeries. Electronic flexible ureteroscope has high resolution of image quality, easy operation, and good durability, with high safety especially for biopsy operations for urinary tumors. Fiber flexible ureteroscope has a thinner lens and lower cost, which can meet the needs of urinary calculi surgery.
Keywords:Electronic flexible ureteroscope  Fiber flexible ureteroscope  Urology