首页 | 官方网站   微博 | 高级检索  
     


Injection transient study using a two-frequency bunch length measurement system at the SSRF
Authors:Li-Wu Duan  Yong-Bin Leng  Ren-Xian Yuan  Zhi-Chu Chen
Affiliation:1. Shanghai Institute of Applied Physics, Chinese Academy of Sciences, Shanghai 201800, China;2. Shanghai Institute of Applied Physics, Chinese Academy of Sciences, Shanghai 201800, China;Shanghai Synchrotron Radiation Facility, Chinese Academy of Sciences, Zhangjiang Campus, Shanghai 201204, China
Abstract:The bunch length can be measured by comparison of two frequency components of a synchrotron beam signal.An online bunch length measurement system has been implemented based on this method.Working frequencies of 3 GHz and 500 MHz were selected,and the raw data was acquired by digital oscilloscope and was resampled and analyzed using the MATLAB software platform at bunch-by-bunch rate.The constructed system was employed to study the bunch length synchronous oscillation phenomenon during injection.The beam experiments demonstrated a time resolution of less than 0.5 ps.
Keywords:Bunch length measurement  Two-frequency method  Bunch-by-bunch  Injection
本文献已被 CNKI 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号