首页 | 官方网站   微博 | 高级检索  
     


Detective quantum efficiency [DQE(0)] of CZT semiconductor detectors for digital radiography
Authors:Giakos  GC Suryanarayanan  S Guntupalli  R Odogba  J Shah  N Vedantham  S Chowdhury  S Mehta  K Sumrain  S Patnekar  N Moholkar  A Kumar  V Endorf  RE
Affiliation:Dept. of Electr. & Comput. Eng., Univ. of Akron, OH, USA;
Abstract:In this paper, the detective quantum efficiency (DQE) of cadmium zinc telluride (CZT) detector samples for digital radiography has been measured. Specifically, this study is aimed at investigating the zero-frequency DQE(0) under different X-ray tube and detector parameters. The experimental results of this study indicate that the DQE(0) of the CZT samples is strongly dependent upon the irradiation geometry. This is attributed to the incomplete charge collection process, which can be further improved by controlling the purity of the samples and the contact type.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23