首页 | 官方网站   微博 | 高级检索  
     


High-accuracy displacement metrology and control using a dual Fabry-Perot cavity with an optical frequency comb generator
Authors:Youichi Bitou
Affiliation:1. Mechanical and Systems Research Laboratories, Industrial Technology Research Institute, Hsinchu, Taiwan;2. Department of Mechanical and Electro-Mechanical Engineering, Tamkang University, No. 151, Ying-chuan Road, Tamsui District, New Taipei City, Taiwan
Abstract:A displacement metrology and control system using an optical frequency comb generator and a dual Fabry-Perot cavity is developed with sub-nm accuracy. The optical frequency comb generator has expanded the displacement measurement range and the dual cavity system has suppressed the environmental fluctuation. We evaluated the absolute uncertainty of the developed displacement measurement system to be approximately 190 pm for the displacement of 14 μm and the accurate displacement control using a phase-locked loop was demonstrated with a resolution of approximately 24 pm.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23