平行定量浓缩仪- 超高效液相色谱- 串联质谱联用快速检测鳗鱼中8 种喹诺酮类药物残留量
引用本文:张宪臣,华洪波,张朋杰,王勇,杨芳,李蓉,胡仪光,卢俊文.平行定量浓缩仪- 超高效液相色谱- 串联质谱联用快速检测鳗鱼中8 种喹诺酮类药物残留量[J].食品科学,2012,33(12):190-195.
作者姓名:张宪臣  华洪波  张朋杰  王勇  杨芳  李蓉  胡仪光  卢俊文
作者单位:1.中山出入境检验检疫局 2.中山市广外附设外语学校
基金项目:广东省自然科学基金项目(7301741);中山市科技计划项目(20103A235);广东检验检疫局科技计划项目(2012GDK49)
摘    要:建立鳗鱼肉中8种奎诺酮残留量的平行定量浓缩仪-超高效液相色谱-串联质谱检测方法。以氘代试剂为内标,样品经酸化乙腈萃取后,平行定量浓缩仪浓缩,用正己烷脱脂,采用超高效液相色谱-串联质谱选择反应监测正离子模式测定,可同时对鱼肉中奎诺酮进行定性和定量。奎诺酮检出限可达0.5μg/kg,定量低限可达1.0μg/kg。在2.5~200ng/mL范围内峰强度与质量浓度的线性关系良好(r>0.99)。方法的平均回收率在81.4%~111.1%之间。方法简单、快速、稳定、准确、灵敏度高,适合于实验室大批量检测水产品样品。

关 键 词:平行定量浓缩仪  超高效液相色谱-串联质谱  奎诺酮  鳗鱼  残留量  

Rapid Simultaneous Determination of Eight Quinolone Residues in Eel by Syncore Analyst-Ultra Performance Liquid Chromatography-Electrospray Ionization Tandem Mass Spectrometry
ZHANG Xian-chen,HUA Hong-bo,ZHANG Peng-jie,WANG Yong,YANG Fang,LI Rong,HU Yi-guang,LU Jun-wen.Rapid Simultaneous Determination of Eight Quinolone Residues in Eel by Syncore Analyst-Ultra Performance Liquid Chromatography-Electrospray Ionization Tandem Mass Spectrometry[J].Food Science,2012,33(12):190-195.
Authors:ZHANG Xian-chen  HUA Hong-bo  ZHANG Peng-jie  WANG Yong  YANG Fang  LI Rong  HU Yi-guang  LU Jun-wen
Affiliation:(1. Zhongshan Entry-Exit Inspection and Quarantine Bureau, Zhongshan 528403, China; 2. Zhongshan Foreign Language School Attached to Guangdong University of Foreign Studies, Zhongshan 528401, China)
Abstract:A method for the simultaneous determination of eight quinolone residues in eel was developed by syncore analyst-ultra performance liquid chromatography-electrospray ionization tandem mass spectrometry(SA-UPLC-MS/MS).Deuterium substituted reagents were added as internal standards before sample extraction with acidified acetonitrile,followed by concentration using a Syncore Analyst and clean-up with hexane.Eight quinolone residues were analyzed by UPLC-MS/MS under positive ion mode.The limit of detection of the method for eight quinolones was 0.5μg/kg,and the limit of quantification was 1.0μg/kg.The calibration curves showed good linearity over the concentration range of 2.5-200 ng/mL(r > 0.99).The average recovery rates were 81.4%-11.1% across three spike levels(n = 6).This method was simple,rapid,stable,reliable,sensitive and suitable for the determination of bulk aquatic products.
Keywords:Syncore Analyst  ultra performance liquid chromatography-electrospray ionization tandem mass spectrometry  qunolones  eel  residues