应用AHP技术自动测量颈动脉内-中膜厚度的临床研究
引用本文:连娟,田瑞霞,陶进绪,祝文虎,徐明才,刘彧,杨艳婷,吴学咏,王璨,任明磊.应用AHP技术自动测量颈动脉内-中膜厚度的临床研究[J].东南国防医药,2010,12(4):293-295,318.
作者姓名:连娟  田瑞霞  陶进绪  祝文虎  徐明才  刘彧  杨艳婷  吴学咏  王璨  任明磊
作者单位:1. 解放军105医院超声科,安徽合肥,230031
2. 中国科技大学电子工程与信息科学系,安徽合肥,230037
3. 上海和睦家医院,上海,200032
基金项目:南京军区医学科学技术创新重点课题 
摘    要:目的探讨超声动脉健康评估软件包(AHP)测量颈动脉内-中膜厚度(CIMT)的临床应用价值。方法应用超声动脉健康评估软件自动测量238名健康人的颈总动脉、窦部和颈内动脉的内-中膜厚度和血管年龄,同时进行手动测量,对不同年龄组的测量值进行比较分析。结果不同部位的CIMT不同(窦部〉颈总动脉〉颈内动脉,P〈0.05)。年龄大于60岁组的CIMT明显高于年龄小于60岁的CIMT(P〈0.05)。随着CIMT的增厚,血管年龄亦增大(P〈0.05)。血管年龄随着年龄增大而增大(P〈0.05),但小于实际年龄(P〈0.05)。自动测量值要明显小于手动测量值(P〈0.05)。结论 AHP能方便准确地测量CIMT,具有应用于临床评估动脉粥样硬化的潜力。

关 键 词:超声  测量  颈动脉  厚度  AHP

Clinical study of automatic measurement of marotid intima-media thickness based on AHP
LIAN Juan,TIAN Rui-xia,TAO Jin-xu,ZHU Wen-hu,XU Ming-cai,LIU Yu,YANG Yan-ting,WU Xue-yong,WANG Can,REN Ming-lei.Clinical study of automatic measurement of marotid intima-media thickness based on AHP[J].Journal of Southeast China National Defence Medical Science,2010,12(4):293-295,318.
Authors:LIAN Juan  TIAN Rui-xia  TAO Jin-xu  ZHU Wen-hu  XU Ming-cai  LIU Yu  YANG Yan-ting  WU Xue-yong  WANG Can  REN Ming-lei
Affiliation:1.Department of Ultrasound,105 Hospital of PLA,Hefei,Anhui 230031,China;2.Department of Electronic Engineering and Information Sciences,University of Science and Technology of China,Hefei,Anhui 230027;3.He Mu-jia Hospital of Shanghai,Shanghai 200032,China )
Abstract:Objective To investigate the clinical value of automatic measurement of carotid intima-media thickness(CIMT) based on AHP.Methods The CIMT and vascular age(VA) of 238 healthy persons were obtained with ultrasound arterial health package.IMT of common carotid artery was measured manually.Results The CIMT of different positions were different(Bulbcommon carotid arteryinternal carotid artery,P〈0.05).The CIMT of the group whose age were older than sixty was significantly higher than that of other groups(P〈0.05).The VA increased with the CIMT thickening(P〈0.05) or the age increasing(P〈0.05).However,the VA was younger than the physical age in all groups(P〈0.05).The automatic measurement values were significantly smaller than the manual ones(P〈0.05).Conclusion The proposed method is accurate and convenient,and has potential to assess atherosclerosis in clinic.
Keywords:AHP