首页 | 官方网站   微博 | 高级检索  
     

奇异值分解滤波联合Frangi滤波对超微血管成像图像质量的影响
引用本文:胡海曼,雷雨蒙,余靖,王琦,雷炳松,叶华容,张舸.奇异值分解滤波联合Frangi滤波对超微血管成像图像质量的影响[J].中国医学影像技术,2024,40(3):430-435.
作者姓名:胡海曼  雷雨蒙  余靖  王琦  雷炳松  叶华容  张舸
作者单位:武汉亚洲心脏病医院心血管内科, 湖北 武汉 430022;华润武钢总医院超声医学科, 湖北 武汉 430080;湖北工业大学电气与电子工程学院, 湖北 武汉 430068
基金项目:湖北省自然科学基金青年项目(2023AFB289)。
摘    要:目的 观察奇异值分解(SVD)滤波联合Frangi滤波对超微血管成像(SMI)图像质量的影响。方法 分别对1具仿体及1例疑诊肝血管瘤患者行CEUS,分别以SVD滤波(A组)、Frangi滤波(B组)、Frangi+SVD滤波(C组)及SVD+Frangi滤波(D组)处理图像,之后比较不同图像的质量。结果 4组仿体及人体肝脏SMI对比组织比(CTR)、对比度噪声比(CNR)及信噪比(SNR)均高于原始CEUS图像。相比原始图像,D组图像质量及分辨率提升最为显著,仿体SMI的 CTR、CNR及SNR分别提升58.04、3.39及48.04 dB,人体肝脏SMI相应参数则分别提升61.85、16.80及49.67 dB,而分辨率分别为原始CEUS图像的1.42倍及1.98倍。结论 SVD滤波联合Frangi滤波可有效提高SMI图像质量。

关 键 词:血管瘤  超声检查  图像质量  奇异值分解  Frangi滤波器
收稿时间:2023/11/21 0:00:00
修稿时间:2023/12/28 0:00:00

Impact of singular value decomposition filter combined with Frangi filter on image quality of superb microvascular imaging
HU Haiman,LEI Yumeng,YU Jing,WANG Qi,LEI Bingsong,YE Huarong,ZHANG Ge.Impact of singular value decomposition filter combined with Frangi filter on image quality of superb microvascular imaging[J].Chinese Journal of Medical Imaging Technology,2024,40(3):430-435.
Authors:HU Haiman  LEI Yumeng  YU Jing  WANG Qi  LEI Bingsong  YE Huarong  ZHANG Ge
Affiliation:Department of Cardiovascular Medicine, Wuhan Asia Heart Hospital, Wuhan 430022, China;Department of Ultrasound Medicine, China Resources & Wisco General Hospital, Wuhan 430080, China;School of Electrical and Electronic Engineering, Hubei University of Technology, Wuhan 430068, China
Abstract:Objective To observe the impact of singular value decomposition (SVD) filter combined with Frangi filter on image quality of superb microvascular imaging (SMI). Methods Contrast-enhanced ultrasound (CEUS) was performed in a phantom and a patient with suspected liver hemangioma. Images in group A were processed with SVD filter, while in group B, C and D with Frangi filter, Frangi+SVD filter and SVD+Frangi filter, respectively, then image quality of SMI were compared. Results Contrast-to-tissue ratio (CTR), contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) of A,B, C and D group were all higher than those of original CEUS images. Compared with the original CEUS images, the image quality indexes and image resolution of group D were most significantly improved, with CTR, CNR and SNR increased by 58.04, 3.39 and 48.04 dB of the phantom, as well as increased by 61.85, 16.80 and 49.67 dB of the patient, respectively. The resolution of the phantom''s and the patient''s SMI images was 1.42 times and 1.98 times of the original CEUS images, respectively. Conclusion SVD filter combined with Frangi filter could effectively improve image quality of SMI.
Keywords:hemangioma  ultrasonography  image quality  singular value decomposition  Frangi filter
点击此处可从《中国医学影像技术》浏览原始摘要信息
点击此处可从《中国医学影像技术》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号