Biodegradable spacer reduces the subacromial pressure: A biomechanical cadaver study
Affiliation:1. Department of Orthopedic Surgery, Physical Medicine and Rehabilitation, University Hospital of Munich, Germany;2. Tel Aviv Sourasky Medical Center, Tel Aviv, Israel;3. Department of Orthopedic Surgery, Wolfson Medical Center, Holon, Israel;4. Advanced Orthopedic Clinic, Assuta Medical Center, Tel Aviv, Israel
Abstract:BackgroundFailure after rotator cuff repair remains a major clinical problem and could be related to excessive pressures from the acromion. Previous studies with irreparable tears showed good clinical results of tendon healing with arthroscopic insertion of a protective biodegradable spacer balloon between the repaired tendon and the acromion. One hypothesis is that compression pressures on the repaired tendon will be reduced by the spacer. This cadaver study aimed to investigate the effects of this subacromial spacer on compression pressures over a repaired supraspinatus tendon in passive motion.MethodsRotator cuff tear and repair were performed in six fresh-frozen cadaveric shoulders, followed by insertion of a biodegradable subacromial spacer. Specimens were tested using a passive shoulder simulator for abduction-adduction, flexion-extension and internal-external rotations. A sensor positioned below the acromion was used to measure compression pressure changes through passive range of motion before and after placement of a subacromial spacer. Peak pressures were measured in adduction-abduction motion, near 90° abduction.FindingsBoth the mean and peak pressures in abduction-adduction were significantly reduced after insertion of the subacromial spacer (from mean 121.7 (SD 9.5) MPa to 51.5 (SD 1.2) MPa and from peak 1749.6 (SD 80.7) MPa to 535.1 (SD 27.6) MPa) (P < 0.0001).InterpretationThe reduced peak pressures and wider load distributions over the sensor during both passive abduction-adduction and flexion-extension motions suggest that the use of the spacer will lead to reduced wear of the repair in patients, and potentially prevent rotator cuff re-tear after surgical repair.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号