1.5T场强下SE、GRE-EPI和弥散加权序列进行脑fMRI的对比研究
引用本文:徐海波,徐丽莹,邓先波,支新,石浩军,邱大胜,王伟,孔祥泉.1.5T场强下SE、GRE-EPI和弥散加权序列进行脑fMRI的对比研究[J].临床放射学杂志,2004,23(5):367-370.
作者姓名:徐海波  徐丽莹  邓先波  支新  石浩军  邱大胜  王伟  孔祥泉
作者单位:430022,武汉,华中科技大学同济医学院协和医院放射科
摘    要:目的 探讨在 1.5T场强下自旋回波 (SE EPI) ,梯度回波 (GRE EPI)和弥散加权序列检测脑活动的能力。资料与方法  5个右利手健康志愿者 ,在 1.5T场强下通过SE EPI、GRE EPI、弥散加权 (b值 =5 0s/mm2 )序列进行复杂对指运动事件相关性脑功能成像的对比研究 ,采用随机自带的fMRI后处理软件获取功能图像。同时对不同序列所获取的功能信号范围、大小面积和功能信号变化幅度进行t检验。结果  5名志愿者在行复杂对指运动时 ,GRE EPI与SE EPI序列在显示神经元活动的分布范围相似 ,但在显示每个激活区面积 ,GRE EPI显示的明显大于SE EPI(t=- 8.195 ,P <0 .0 1)。GRE EPI和SE EPI序列均显示在刺激与静息间存在明显差异 (tgre=2 .788,Pgre<0 .0 1;tse=- 2 3.2 88,Pse<0 .0 1) ,且两种序列显示的功能信号变化幅度有显著差异 (t=- 9.2 85 ,P <0 .0 1)。而弥散加权序列未能检测出脑功能活动信号。结论 在 1.5T场强下现有的SE或GRE EPI序列均可获取脑功能活动信号 ,GRE EPI序列比SE EPI反映活动神经元敏感性高 ,主要是采集到引流大静脉的信号

关 键 词:  对指运动  磁共振功能成像  梯度回波  自旋回波

A Comparative Study of Brain BOLD Fmri with SE-EPI, GRE-EPI and DWI Sequences at 1.5 Tesla
XU Haibo,XU Liying,DENG Xianbo,et al..A Comparative Study of Brain BOLD Fmri with SE-EPI, GRE-EPI and DWI Sequences at 1.5 Tesla[J].Journal of Clinical Radiology,2004,23(5):367-370.
Authors:XU Haibo  XU Liying  DENG Xianbo  
Affiliation:XU Haibo,XU Liying,DENG Xianbo,et al. Department of Radiology,The Affiliated Union Hospital,Tongji Medical College,Huazhong University of Science and Technology,Wuhan,Hubei Province 430022,P. R. China
Abstract:Objective To evaluate SE-EPI, GRE-EPI and DWI sequences at 1.5 tesla in estimating human brain activities. Materials and Methods Brain BOLD fMRI with SE-EPI, GRE-EPI and DWI (b=50 s/mm 2) sequences at 1.5 tesla was carried out in 5 right-handed healthy volunteers who was performing complex finger-tapping movement during the exam. The functional brain activation was automatically mapped by using the postprocessing software the MR unit carried. The functional signal extent, size and area, as well as the fluctuation range of the signal obtained from different sequences were compared with each other and analyzed by using students' group t test. Results During complex finger-tapping movement, GRE-EPI showed quite the same distribution extent of neuronal activation as SE-EPI did, although the activated area displayed on GRE-EPI was much larger than that displayed on SE-EPI (t=8.195, P<0.01). Significant difference in signal intensity existed between the rest phase and the activation phase on both GRE-EPI (t=2.788, P<0.01) and SE-EPI (t=-23.288, P<0.01). Significant difference in signal fluctuation was also found between SE-EPI and GRE-EPI, with t=-9.285 and P<0.01. DWI could not demonstrate the signal of brain activity. Conclusion At 1.5 tesla, event-related brain activation signal can be documented by both SE-EPI and GRE-EPI BOLD fMRI. GRE-EPI is more sensitive than SE-EPI in detecting neuronal activation as GRE-EPI can get the signal from the large draining vein.
Keywords:Brain  Finger tapping  Functional  Mangetic resonance imaging