比色板与电脑比色仪比色准确性的分析
引用本文:徐明明,刘峰,张峰,丁中.比色板与电脑比色仪比色准确性的分析[J].中华口腔医学杂志,2008,43(10).
作者姓名:徐明明  刘峰  张峰  丁中
作者单位:1. 北京大学口腔医学院·口腔医院特诊科,100081
2. 北京大学口腔医学院·口腔医院第一门诊部,100034
3. 北京大学口腔医学院·口腔医院义齿加工中心,100034
摘    要:目的 通过分析比色板与电脑比色仪比色的准确性,探讨电脑比色仪对比色板色标未能覆盖颜色的精确表达作用.方法 选取天然牙颜色分布相对均一的单颗前牙全瓷冠修复患者120例,按照随机数字表分为两组(V组和S组),每组60例.V组采用比色板,S组采用电脑比色仪,对患者的天然牙进行比色;再根据测色结果分为色标颜色组(V1、S1)和中间颜色组(V2、S2).修复后使用电脑比色仪对所有修复体及V组比色参照的天然牙进行测色,分别计算每组修复体中部与天然牙中部的总色差△E*ab,对结果进行配对t检验.结果 V组△E*ab(3.92±1.59)与S组△E*ab(2.23±0.96)的差异有统计学意义(P<0.01);V1组△E*ab(2.29±0.88)与V2组△E*ab(4.42±1.42)的差异有统计学意义(P<0.01);S1组△E*ab(2.12±0.84)与S2组△E*ab(2.27±1.01)的差异无统计意义(P=0.014).V2组AE*ab与S2组△E*ab的差异有统计学意义(P<0.01);V1组△E*ab与S1组△E*ab的差异无统计学意义(P=0.698).结论 对于比色板色标未能覆盖的颜色,电脑比色仪的定位与表达比单纯用比色板比色更准确,利于指导修复体的颜色再现.

关 键 词:美学  牙科  比色  电脑比色仪  比色板

Comparison of accuracy between visual and spectrophotometric shade matching
XU Ming-ming,LIU Feng,ZHANG Feng,DING Zhong.Comparison of accuracy between visual and spectrophotometric shade matching[J].Chinese Journal of Stomatology,2008,43(10).
Authors:XU Ming-ming  LIU Feng  ZHANG Feng  DING Zhong
Abstract:Objective To compare visual and spectrophotometric shade matching using Vita Easyshade and to evaluate its accuracy.Methods One hundred and twenty participants with single anterior full ceramic restorations were included in this study. Subjects were randomly assigned to 2 groups,the visual shade matching group (Group V) and the spectrophotometric shade matching (Vita Easyshade) group(Group S).In each group, subjects were allocated into 2 subgroups,those matched with shade tabs (V1,S1) and those out of shade tabs (V2,S2).Following the treatment,the Commission Internationale de I'Eclairage lab parameters were tested using the spectrophotometric,and the AE*ab between the restoration and contrast teeth was calculated.Results There were significant differences between AE*abv (3.92±1.59) and △E*abs (2.23±0.96),and also between AE*abv2(4.42±1.42) and AE*abs2(2.27±1.01),AE*abv1(2.29±0.88) and △E*abv2(P<0.01).There was no significant difference between AE*abv1 and AE*abs1 (2.12±0.84) (P=0.698),nor between AE*abs1 and AE*abs2 (P=0.014).Conclusions Spectropotometric (Vita Easyshade) shade matching method is more accurate than visual shade matching when used on tooth that has out-of-shade-tab eolour expression.It provides better colour reappearance,than does visual colour matching.
Keywords:Esthetics  dental  Shade selection  Computer-aided spectrophotometer  Shadeguide