首页 | 官方网站   微博 | 高级检索  
     


Serum hCG assay: A method for detection of contamination of fetal blood samples
Authors:M Dommergues  V Bunduki  F Muller  L Mandelbrot  N Morichon-Delvallez  Y Dumez
Affiliation:1. Malernité Port-Royal, Paris, France;2. Service de Biochimie, Hôpital Ambroise Paré, Boulogne, France;3. Laboratoire de Cytogénétique Prénatale, Hôpital Necker-Enfants Malades, Paris, France
Abstract:Serum human chorionic gonadotrophin (hCG) can be assayed in specimens obtained by percutaneous fetal blood sampling to check for the absence of maternal blood or amniotic fluid contamination. In order to assess the accuracy of this approach, we measured serum hCG in 44 pure fetal blood samples obtained by intracardiac puncture. The mean fetal serum hCG concentration was 52 IU/l, and the ratio of maternal to fetal serum hCG concentration never exceeded 1 · 1 per cent, which represents the smallest contamination rate detectable by this method.
Keywords:hCG  Fetal blood sampling
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号